Section 2-3 - Facility Needs
2-3-2
Mandatory Site Requirements
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in
Section 2-3-3 "Site Recommendations"
.
•
A dedicated single branch power outlet of adequate amperage (see
Table 2-3 on page 2-2
) that
meets all local and national codes and is located less than 2.5 m (8.2 ft) from the unit’s proposed
location. Refer to:
Section 2-2-2 "Electrical Requirements" on page 2-2
.
•
A door opening of at least 76 cm (2.5 ft) in width (if using the Cart).
•
The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
•
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•
Power outlets for other medical equipment and gel warmer.
•
Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
•
Clean and protected space for storage of probes (either in their case or on a rack).
•
Material to safely clean probes (performed using a plastic container, never metal).
•
In the case of a network option:
-
An active network outlet in the vicinity of the ultrasound unit.
-
A network cable of appropriate length (regular Pin-to-Pin network cable).
-
An IT administrator who will assist in configuring the unit to work with your local network. A
fixed IP address is required. Refer to the form provided in
Figure 3-141 on page 3-120
for
network details that are required.
NOTE:
All relevant preliminary network port installations at the prepared site must be performed by authorized
contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing
of the equipment.
2-3-3
Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements
section, above.
•
Door opening of 92 cm (3 ft) in width (if using the Cart).
•
Accessible circuit breaker for a dedicated power outlet.
•
Sink with hot and cold running water.
•
Receptacle for bio–hazardous waste, for example, used probe sheaths.
•
Emergency oxygen supply.
•
Storage area for linens and equipment.
•
Nearby waiting room, lavatory, and dressing room.
•
Dual level lighting (bright and dim).
•
Lockable cabinet for software and manuals.