Section 5-2 - General Information
5-2-4-1-1
Special B Mode Techniques
a.)
HI
(Coded Harmonic Imaging)
In one method of HI the RX-Frequency is doubled, so that the radial resolution is increased
due to the higher RX-Frequency.
The second method of HI is pulse-inversion: 2 TX-Beams are shot to the same Tissue-location,
one with positive, one with negative polarity. The subtraction of both shots (Subtraction Filter)
brings to bear the nonlinear-echo-reflection-properties of the tissue, which is very useful with
extremely difficult-to-image patients.
5-2-4-2
M-Mode
1.) RBI
see:
5-2-4-1 B-Mode
a.) DMA section
B-mode-Data from RBI is written via SP0 into SDRAM Fifo Buffer memory. DMA Controller 1
transfers the data into PC main memory where scan conversion is performed per software,
i.e. the sweep image is generated (scaling and interpolation between lines).
CineMode: CineMode-Memory is the PC main memory.
2.) RBF
see:
5-2-4-1 B-Mode
5-2-4-3
D-Mode (Pulsed Wave Doppler)
1.) RBI
-
PRF-generator; see:
5-2-4-1 B-Mode
-
After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum,
where the LowPassFilter cuts HF.
Mixer and Magnitude-Calculator arrange Complex Demodulation.
a.) DMA section
D-mode-Data from RBI is written via SP1 into SDRAM Fifo Buffer memory. DMA Controller 1
transfers the data into PC main memory where scan conversion is performed per software,
i.e. the sweep image is generated (scaling and interpolation between lines).
CineMode: CineMode-Memory is the PC main memory.
2.) RBF
see:
5-2-4-1 B-Mode
Содержание H48651KR
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO ...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents ...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index ...
Страница 515: ......