Chapter 5 - Components and Functions (Theory)
5-55
5-13-2-6
Delete all Patients
1.) Click the DELETE ALL PATIENTS... button. Following WARNING message appears on the screen.
5-13-2-7
Export System Data
Select the EXPORT SYSTEM DATA button on the “Service Tools” page to Full Backup the System
State. This includes dump-files and text files, the full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
Operation see:
Section 7-4-2 "Export Log’s and System Data" on page 7-9
.
5-13-2-8
Battery Calibration
Select the BATTERY CALIBRATION button to perform fuel gauge recalibration of the battery.
Operation see:
3-6-4-2-1 "Recalibration by means of Internal Battery Charger" on page 3-77
.
5-13-2-9
Keyboard Layout
To change the keyboard layout to different languages.
Operation see:
Section 6-9 "Modification of Keyboard Layout" on page 6-10
.
NOTE:
Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
see:
Section 8-7 "Replacement of Key Caps (by special native language keys)" on page 8-25
.
5-13-2-10
BootUp Screen
With this button you can give in your own name. You can see it, when you Bootup the Voluson i /
Voluson e.
5-13-2-11
Monitor Test
Select the MONITOR TEST button to perform color calibration.
Operation see:
Section 6-3 "LCD Display (Monitor) Adjustment" on page 6-2
.
5-13-2-12
Printer
Installation of printers or editing their settings is possible without entering the Windows Desktop.
Operation see:
3-7-4 "Printer Installation manually" on page 3-87
.
5-13-2-13
Probe EEPROM
NOTE:
“Probe EEPROM” is available from software 8.2.2 and software CC200 BT’14 and higher.
Figure 5-30 Warning message
WARNING
WARNING
!! WARNING:
If you select the YES button, all patients data, studies, images and measure report data
will be deleted permanently from the hard disk and cannot be recovered!
WARNING
WARNING
!! WARNING:
Only accessories explicitly recognized and released by the system manufacturer GE
may be used in connection with the system.
NOTICE
It is recommended to connect and re-program one probe after the other.