8-74
Section 8-20 - Replacement of the Voluson Station Components
8-20-14
Replacement of the Top Column Back Cover
8-20-14-1
Manpower
One person, 15 minutes
8-20-14-2
Tools
Phillips screwdriver 1
8-20-14-3
Preparations
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Remove the Top Shelf Additional Bottom Cover as described in
Section 8-20-12 on page 8-72
or
remove the optional Probe Mux Bos as described in
Section 8-20-9 on page 8-68
.
3.) Remove the Top Shelf Bottom Rear Cover as described in
Section 8-20-13 on page 8-73
.
8-20-14-4
Top Column Back Cover - Removal Procedure
1.) Screw off the 4 screws and disconnect the power cable.
2.) Carefully remove the Top Column Back Cover.
8-20-14-5
Top Column Back Cover - Installation Procedure
1.) Place the new Top Column Back Cover, connect the power cable and fasten it with the 4 screws.
2.) Mount the Top Shelf Bottom Rear Cover as described in
Section 8-20-13-5 on page 8-73
.
3.) Mount the Top Shelf Additional Bottom Cover as described in
Section 8-20-12-5 on page 8-72
or
mount the optional Probe Mux Box as described in
Section 8-20-9-5 on page 8-69
.
Figure 8-97 unscrew the 4 screws / disconnect the power cable