Section 7-4 - Screen Captures and Logs
7-4-2-2
Export Log’s and System Data (via Service Page)
1.) Press the
UTILITIES
key on the keyboard, select the SYSTEM item from the menu area.
2.) Select the
SERVICE
page on the screen. The “password window” appears automatically.
3.) Enter the password
SHE
and click the ACCEPT button to display the Service Tools window.
4.) Click on the EXPORT SYSTEM DATA... button to Full Backup the System State. This includes
dump files (see:
Section 7-4-2-2-1
) and text files, full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
7-4-2-2-1
Dump-file
Every time an error message like
Figure 7-8
is produced, a dump-file and a text file containing the
error dump and the error message are created in D:\export. Up to 20 dump files are stored there.
After clicking on OK the system reboots automatically.
Figure 7-7 Service Tools window
Figure 7-8 System has encountered a problem
with Software 6.x.x and 7.x.x
Export System Data
Export System Data
with Software 8.x.x