background image

GE Healthcare
CARESCAPE Monitor B650
Technical Manual

Software Version 2
Hardware Version B650-01 

All specifications subject to change without notice.

 

English
 

2062973-002 paper

19 September 2016

© 2013, 2016 General Electric Company. All rights reserved.

Содержание CARESCAPE Monitor B650

Страница 1: ... Monitor B650 Technical Manual Software Version 2 Hardware Version B650 01 All specifications subject to change without notice English 2062973 002 paper 19 September 2016 2013 2016 General Electric Company All rights reserved ...

Страница 2: ...mation in this manual applies to the software and hardware versions listed on the first page of this manual Due to continuing product innovation specifications in this manual are subject to change without notice ...

Страница 3: ...ols 6 3 System overview 9 3 1 System introduction 9 3 2 System components 9 3 2 1 Monitor 9 3 2 2 Software 10 3 2 3 Input devices 10 3 2 4 Acquisition modules 12 3 2 5 CARESCAPE Network MC or S 5 Network 12 3 2 6 CARESCAPE Network IX 13 3 2 7 Unity Network ID connectivity device 14 3 2 8 Secondary display 14 3 2 9 Printers and recorders 15 3 2 10 Service Interface 16 3 3 Controls and connectors 17...

Страница 4: ... environmental requirements 54 6 Hardware installation 57 6 1 Installing a battery into the patient monitor and the PDM module 57 6 1 1 Testing the battery charge 57 6 1 2 Installing the battery into the patient monitor 58 6 1 3 Installing the battery into the PDM module 58 6 2 Mounting the patient monitor 59 6 3 Connecting a secondary display 60 6 3 1 Connections to D15K and D19KT displays 60 6 4...

Страница 5: ... 7 12 Setting the host asset number 87 7 13 Changing passwords 87 7 14 Restarting the patient monitor 88 7 15 Setting up the remote service 88 7 15 1 Configuring the remote service 88 7 15 2 Enabling remote service agent connection 89 7 16 Transferring settings from one patient monitor to other patient monitors 90 7 16 1 Saving settings 90 7 16 2 Loading settings 91 7 16 3 Activating settings 91 7...

Страница 6: ... 11 IX printers 114 8 3 12 Insite with EXC 114 8 3 13 Test completion 114 9 Theory of operation 115 9 1 Main components 116 9 1 1 Power management subsystem 116 9 1 2 CPU subsystem 118 9 1 3 Display subsystem 120 9 1 4 User interface subsystem 121 9 1 5 External Interfaces 123 9 1 6 Pivoting module frame 125 10 Maintenance and checkout 129 10 1 Visual inspection 130 10 2 Electrical safety checks 1...

Страница 7: ... 11 7 Error messages and codes 154 11 8 Problems and solutions 158 11 8 1 Start up failures 158 11 8 2 User interface issues 161 11 8 3 Incorrect system time 163 11 8 4 License issues 164 11 8 5 Recorder issues 165 11 8 6 Acquisition module problems 165 11 8 7 CARESCAPE Network communication issues 168 11 8 8 S 5 Network communication issues 172 12 Disassembly and reassembly 175 12 1 Disassembly g...

Страница 8: ...r unit assembly 197 12 3 11Detaching the Base unit 198 12 4 Front Unit Assembly 201 12 4 1 Detaching the Front Unit Assembly 201 12 4 2 Detaching the LCD Display Unit 203 12 4 3 Replacing the User Interface Board 205 12 4 4 Replacing the Trim Knob and Trim Knob Encoder 205 12 4 5 Replacing the Keypad 206 12 4 6 Replacing the Front Unit Assembly FRU 207 12 4 7 Replacing the WLAN Assembly FRU 208 13...

Страница 9: ...or the instructions necessary to operate the device safely in accordance with its function and intended use 1 2 Intended audience of the manual This manual is intended for service representatives and technical personnel who install maintain troubleshoot or repair this device 1 3 Conventions used in this manual Within this manual special styles and formats are used to distinguish between terms view...

Страница 10: ...nd case menu in OR and PACU software 1 4 Illustrations and names This manual uses illustrations as examples only Illustrations in this manual may not necessarily reflect all system settings features configurations or displayed data Names of persons institutions and places and related information are fictitious any similarity to actual persons entities or places is purely coincidental 1 5 Ordering ...

Страница 11: ...mo Corporation PiCCO is a trademark of Pulsion Medical Systems SE WMM WPA and WPA2 are trademarks of Wi Fi Alliance Covidien BISx Bispectral Index BIS and Nellcor are trademarks of a Medtronic company 1 8 Responsibility of the manufacturer GE is responsible for the effects on safety reliability and performance of the equipment only if Assembly operations extensions readjustments modifications serv...

Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...

Страница 13: ...ctions The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system Consideration relating to the choice shall include use of the accessory in the patient vicinity and evidence that the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601 1 and or I...

Страница 14: ... instructions DANGER Shock hazard Dangerous voltage To reduce the risk of electric shock do not remove cover Refer servicing to qualified service personnel This symbol is identified by a yellow background black triangular band and a black symbol Electrostatic sensitive device Connections should not be made to this device unless ESD precautionary procedures are followed Non ionizing electromagnetic...

Страница 15: ...ation to the patient including direct cardiac application Type CF IEC 60601 1 defibrillator proof protection against electric shock Isolated floating applied part suitable for intentional external and internal application to the patient including direct cardiac application Safety ground Remove power cord from the mains source by grasping the plug Do not pull on the cable ...

Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...

Страница 17: ...faces The power management subsystem provides the operating voltages for the electronics of the device and takes care of the battery management It consists of an AC DC power supply a DC DC power management board and an optional lithium ion battery The CPU board is the main board of the monitor It takes care of the user input and acquisition data processing and displays the processed information on...

Страница 18: ...ing to a thermal paper Both options also are available as field upgrades 3 2 2 Software The patient monitor is highly configurable and provides many monitoring possibilities with a flexible software licensing model The monitor supports care area specific software packages for OR PACU ICU ED and NICU Each dedicated software package provides a comprehensive feature set for the different monitoring n...

Страница 19: ...ted to the monitor or display via one of the USB connectors The mouse allows you to select any on screen items without a Trim Knob control or a touchscreen display Remote control The remote control provides all patient monitor controls on a portable component with a Trim Knob control The remote control is connected to the patient monitor via one of the USB connectors Barcode reader The barcode rea...

Страница 20: ...h or one double width E modules at a time Refer to the patient monitor s supplemental information manual for a list of compatible acquisition devices and to the patient monitor s user s manual for a list of parameters each module measures 3 2 5 CARESCAPE Network MC or S 5 Network The patient monitor is compatible both with the CARESCAPE Network MC and the S 5 Network infrastructures The optional W...

Страница 21: ...ion and allows patient data to be sent to an iCentral central station 3 2 6 CARESCAPE Network IX The patient monitor may be connected to the CARESCAPE Network IX The IX Network provides you access for example to the MUSE server for MUSE 12SL reports and to the IX printers It also enables centralized Webmin access for service personnel from within the hospital and the InSite with ExC remote service...

Страница 22: ... Network Interface Device ID Operator s Manual and the patient monitor s supplemental information manual for a list of compatible peripheral devices and to the patient monitor s user s manual for the peripheral device parameter data displayed on the patient monitor 3 2 8 Secondary display The patient monitor supports a secondary clone display that is capable of displaying the same image as the int...

Страница 23: ...tient monitor s user s manual for more information about printing The patient monitor supports printing to a laser printer that is connected to the patient monitor via the IX Network or directly to the IX connector in the patient monitor to a laser printer that is connected to a CIC Pro Clinical Information Center on the MC Network to a laser printer that is connected to an iCentral on the S 5 Net...

Страница 24: ...he other way to access Webmin locally is from a configured service PC that is connected to the patient monitor with an Ethernet crossover cable Remote access to Webmin The user also can access Webmin remotely using a configured service PC over the IX Network InSite with ExC InSite with ExC provides a set of software applications to manage diagnose and track systems at customer sites by using the I...

Страница 25: ...touch screen option Alarm light Power indicators 1 Alarm light 2 Power on standby button 3 Battery power mains power indicators 4 Abbreviated integrated keypad 5 Trim Knob control 6 Ambient light detector lens 7 Audio alarm paused off area blue 8 Alarm light area blue yellow or red 6 7 9 10 11 8 ...

Страница 26: ...tor battery is charging and flashing in case of battery failure or missing battery 3 3 2 Side views 1 Battery cover lock 2 Battery cover 3 Module slot for one double width or two single width modules 4 Defibrillator ECG and IABP synchronization E modules only 5 Release switch for the pivoting module frame 6 Recorder optional NOTE The Defibrillation synchronization connector can be used only with E...

Страница 27: ...for troubleshooting 2 Cable clamp for power cord 3 Equipotential connector 4 Receptacle for power cord and fuse holder 5 Device label 6 Slide mount with connector for PDM 7 Slide mount with connector for PSM 8 Advanced Interface Board or Standard Interface Board ...

Страница 28: ...ice information 3 4 1 Service requirements Follow the service requirements listed below Refer equipment servicing to GE authorized service personnel only Any unauthorized attempt to repair equipment under warranty voids that warranty It is the user s responsibility to report the need for service to GE or to one of their authorized agents Failure on the part of the responsible individual hospital o...

Страница 29: ... the information on a device label is shown below The product code for CARESCAPE Monitor B650 01 is SEW 3 5 Product security The patient monitoring software incorporates an assortment of security features designed to allow a flexible approach to safe and secure implementation focusing on the principles of confidentiality integrity and availability These features assist you in using the system in a...

Страница 30: ...tient monitor contains features that will help implement and enforce an organization s method Both clinical and Webmin applications have an authorization mechanism to provide information to the user Audit The ability to record and examine system activity is crucial to a successful information security program as well as a regulatory requirement in most environments The patient monitor stores syste...

Страница 31: ...n to the organization s general purpose computing network Traffic between the organization s network and IX port of the patient monitor must be limited to the following packet flows listed below Inbound Packets that are part of the communication initiated by authorized devices in the organization s network are allowed to go out of the IX Network reflexive Outbound Packets that are part of the comm...

Страница 32: ...he wireless network The wireless MC Network shall be logically separated to its own virtual LAN with a dedicated SSID The S 5 Wireless Network shall have dedicated access points that are physically isolated from all other networks Security Authentication and Confidentiality The data transmitted on the wireless network can be secured by one of the following methods WEP 64 bit NOTE S 5 only WEP 128 ...

Страница 33: ...se Do not reuse Battery monitor The flashing orange symbol indicates that there is a battery failure missing battery Battery monitor The solid orange symbol indicates that the battery is being charged Battery monitor The solid green symbol indicates that the monitor is being used on battery power Battery monitor The battery slot cover is open closed Battery monitor Test button on the battery to ch...

Страница 34: ...tery PDM Communication PDM Power indicator PDM On standby button Standby or power indicator Signal power input Signal power output Signal power input output combined ON Power connection to the mains Power supply connector Power switch The following symbols appear on one or more of the devices ...

Страница 35: ... for digital or analog source Color display connector port Color video input Video input connector for digital or analog source Color video output digital Video output for analog source Color video output Video output for digital source USB connector Ethernet connectors The following symbols appear on one or more of the devices ...

Страница 36: ...armful ingress of water Components not marked with an IPXn code are rated as Ordinary no protection against fluid ingress All other IPXn rated components have the degree of protection per the n rating Latex free D lite Pedi lte Add date Home Return to the main display Alternating current Green symbol on the B650 and B450 monitor front panel the monitor is being used on mains power Direct current T...

Страница 37: ...unded to the AC power source Defibrillator synchronization connectors WLAN Class 2 Identifier Stacking limit by number Date of manufacture This symbol indicates the date of manufacture of this device The first four digits identify the year and the last two digits identify the month The following symbols appear on one or more of the devices ...

Страница 38: ...ufacturer name and address Batch or lot number Abbreviation for label part number Catalogue or orderable part number Device serial number Device hardware version Mass of typical portable RGM respiratory gas monitor configuration The indicated mass 12 kg in this example varies per RGM configuration Atmospheric pressure limitations Temperature limitations Humidity limitations Keep dry Protect from r...

Страница 39: ...may contain mercury which must be recycled or disposed of in accordance with local state or country laws Within this system the backlight lamps in the patient monitor display contain mercury European authorized representative European Union Declaration of Conformity Indicates that the product is certified for both the U S and Canadian markets to the applicable U S and Canadian standards FCC USA on...

Страница 40: ...use of such electronic information products will not result in any severe environmental pollution any bodily injury or damage to any assets The unit of the period is Year In order to maintain the declared EFUP the product shall be operated normally according to the instructions and environmental conditions as defined in the product manual and periodic maintenance schedules specified in Product Mai...

Страница 41: ...er Japan TELEC requirements Brazil only Approved under ANATEL Agência Nacional de Telecomunicações requirements South Africa only Approved under ICASA Independent Communications Authority of South Africa requirements Korea only Approved under KCC Korea Communications Commission requirements The following symbols appear in the software user interface Alarm off indicator Displays in the upper right ...

Страница 42: ...ild 2 0 7 or earlier Configuration warning Displays when the priority setting for HR PR high low or SpO2 low has been set to low Check the alarm configuration Displays in the upper left corner of the alarm area With software version 2 host build 2 0 7 or earlier General warning sign Displays when the priority setting for HR PR high low or SpO2 low has been set to low Displays in the Lower part of ...

Страница 43: ...ection indicator Indicates the monitor is connected to the Wireless Local Area Network WLAN Network WLAN signal strength The number of segments corresponds to the signal strength four segments indicate strong signal one segment weak signal Monitor battery is full Monitor battery green The higher the charge the bigger the green bar within the symbol Numbers indicate the remaining run time Monitor b...

Страница 44: ... has an associated snapshot Beat volume icon Adjust the volume of the QRS beep tone Also the beat source indicator Displays next to the selected beat source Respiration indicator Indicates a breath is detected by the impedance respiration algorithm BIS and Entropy sensor impedance check indicator gray Displays for each sensor as the impedance check is in progress BIS and Entropy sensor impedance c...

Страница 45: ...nt of time the SpO2 saturation is outside the limits before alarms are generated SpO2 signal strength indicator Indicates the signal strength with three asterisks indicating the strongest signal NMT Stimulus beep volume icon Adjust the volume of the tone that sounds when a stimulus pulse is generated Progress bar Indicates the amount of time remaining until the next automatic measurement The follo...

Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...

Страница 47: ...thernet crossover cable Depending on the monitor hardware configuration the crossover cable is connected either to the IX connector or to the MC connector Remote Webmin You can access Webmin remotely from a configured service PC that is connected to the patient monitor over the IX Network Requirements for service PC Network board with Ethernet port and TCP IP network installed Internet Explorer V6...

Страница 48: ...rom a live IX Network when a patient is admitted you will temporarily lose the services provided by the IX Network e g access to the IX printers and MUSE reports WARNING Non medical equipment does not provide the same level of protection against electrical shock Do not touch the patient and any part of non medical equipment at the same time Some examples of non medical equipment are laser printers...

Страница 49: ...ation on how to configure the IP address refer to the PC s documentation 5 Launch a web browser on the service PC 6 In the Address field type https IX IP address 10000 and press Enter NOTE IX IP address is the IX Network IP address of the patient monitor The Login to Webmin dialog box displays 7 Continue to 4 5 Login to Webmin Closing Webmin 1 Restore the patient monitor s original IX Network conf...

Страница 50: ...mode Refer to Closing Webmin on page 43 1 Connect a service PC to the MC connector on the patient monitor using a crossover cable 2 Select Monitor Setup Service calibrations 3 Log in with your biomed username and password and press Enter 4 Record the MC IP Address shown next to the Start Maintenance Mode menu button MC IP Address ______________________ MC Netmask ______________________ 5 Select St...

Страница 51: ... PC 3 Disconnect the crossover cable from the monitor and from the service PC 4 Reconnect the monitor back to the S 5 or MC Network if applicable 4 4 Remote access to Webmin using a service PC over the IX Network NOTE This connection method is supported only if the monitor is equipped with the Advanced Interface Board that provides both the IX and the MC connectors and the monitor is connected to ...

Страница 52: ...the service PC 6 In the Address field type https IX IP address 10000 and press Enter NOTE IX IP address is the IX Network IP address of the patient monitor The Login to Webmin dialog box displays 7 Continue to 4 5 Login to Webmin Closing Webmin 1 Restore the original network settings in the service PC 2 Disconnect the service PC from the live IX Network 4 5 Login to Webmin 1 In the Login to Webmin...

Страница 53: ...t password for the username biomed Refer to section 7 13 Changing passwords for details on how to change the default password The Webmin application opens and defaults to the Information tab Figure 1 Webmin user interface when accessed using the integrated browser Figure 2 Webmin user interface when accessed using the service PC ...

Страница 54: ... keyboard locale setting for the alphanumeric keyboard and the barcode reader Licenses to enable and activate a software package to enable and activate software features and to upload and activate a license file MUSE 12SL to configure the host for sending and viewing 12SL information Modules to configure some acquisition module specific settings after corrective maintenance or for administrative p...

Страница 55: ...and for printing a test page Remote Service to configure and control the Insite with Exc remote service tool Restart to shutdown and restart the patient monitor automatically via Webmin NOTE The patient must be discharged in order to enable monitor restart via Webmin Settings to transfer platform and or clinical settings from one patient monitor to another to take backup copies of the settings to ...

Страница 56: ...tion of the patient monitor and the connected peripheral devices The Device Information module shows the hardware and software information of the patient monitor and the connected peripheral devices 4 8 Webmin diagnostics modules Access Webmin service interface to view hardware statistics ping a network device view WLAN diagnostics and view or download log files The Hardware Statistics module disp...

Страница 57: ...The patient monitor collects information about different system events and errors to log files These log files help troubleshooting problems in the patient monitor and the connected peripheral devices WLAN diagnostics Use this Webmin feature for troubleshooting WLAN related problems This Webmin module provides information about the WLAN driver WLAN status and the detected access points ...

Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...

Страница 59: ... a list of compatible network and bedside devices Refer to the patient monitor s supplemental information manual for a list of compatible supplies and accessories Refer to the patient monitor s supplemental information manual and Unity Network Interface Device ID Operator s Manual to see compatible peripheral devices WARNING BEFORE INSTALLATION Compatibility is critical to safe and effective use o...

Страница 60: ...work 5 3 2 Wireless MC Network The wireless MC Network infrastructure shall be installed according to the CARESCAPE Wireless Network Configuration Guide Ensure that the wireless coverage area is adequate for the installation Refer to the sections 7 2 2 Selecting and configuring CARESCAPE Network 7 2 4 Configuring WLAN and 7 4 Setting unit and bed name to see the configuration information you need ...

Страница 61: ...installed prior to the installation of the patient monitor Mounting hardware for the patient monitor either for a stand alone installation or for an installation to an anesthesia machine or to a ventilator Mounting hardware for the PSM module Mounting hardware for the PDM module Mounting hardware for the displays Mounting hardware for the Unity Network ID connectivity device NOTE Refer to the CARE...

Страница 62: ...ion which affords sufficient ventilation The ventilation openings of the device must not be obstructed EMI RFI interference WARNING Do not use the monitor in high electromagnetic fields for example during magnetic resonance imaging WARNING Other equipment may interfere with the system even if that other equipment complies with CISPR emission requirements CAUTION EMC Magnetic and electrical fields ...

Страница 63: ...ment near the system may cause unexpected or adverse operation of this device system Consult qualified personnel regarding device system configuration The patient monitor should be isolated from sources of strong electromagnetic and radio frequency interference NOTE Refer to the patient monitor s supplemental information manual for more information ...

Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...

Страница 65: ...module housing or frames 6 1 Installing a battery into the patient monitor and the PDM module The batteries for the patient monitor and the PDM module if included are shipped separately and need to be installed and fully charged prior to taking into use WARNING PHYSICAL INJURY Make sure the battery is completely inserted and the battery door is completely closed Falling batteries could seriously o...

Страница 66: ...lly charged 6 1 2 Installing the battery into the patient monitor 1 Open the battery cover by turning the battery cover lock 90 degrees clockwise 2 Insert the battery with the test indicator side up and the connector end first all the way into the battery slot To remove the battery pull it out from the cord 3 Push the cover back up and lock it in place by turning the lock 90 degrees counter clockw...

Страница 67: ... compatible mounting hardware for the patient monitor Install the patient monitor to the mounting hardware according to the installation instructions included with the mounting hardware WARNING Never install equipment above the patient WARNING Use only manufacturer specified mounts WARNING To prevent liquids from entering the monitor do not tilt the monitor more than 15 degrees CAUTION The device ...

Страница 68: ...djustments NOTE All installations should be compliant with IEC 60601 1 1 and local electrical codes NOTE The patient monitor with a non medical grade display which is IEC 60950 rated or equivalent meets UL and IEC specifications if a medical grade isolation is used If a non medical grade display is to be used the configuration must meet the IEC 60601 1 1 standard Refer to IEC 60601 1 1 for require...

Страница 69: ...ard USB A to USB B cable up to 5 meters In the following example two 5 meter USB extenders plus a standard 5 meter USB cable extend the remote display up to 15 meters from the patient monitor USB connection 1 Connect the Type A plug of the 1st USB extender to one of the downstream ports Type A USB port on the back of the patient monitor 2 Connect the Type A plug of the 2nd USB extender to the Type...

Страница 70: ... a PSM or a PDM module You can install a PSM or a PDM module either to the patient monitor or to a remote mounting solution Connecting a PSM or a PDM module to the patient monitor 1 Connect a module by aligning it with the insertion guides on the pivoting module frame 2 Push the module into the frame until it clicks and stops Installing a PSM module to a pole mount Ensure that the Pole Mount for P...

Страница 71: ...iously or fatally injure neonatal or other vulnerable patients WARNING PHYSICAL INJURY Do not install the PDM above a patient Leaks from the battery cells can occur under extreme conditions The liquid is caustic to the eyes and skin If the liquid comes in contact with eyes or skin flush with clean water and seek medical attention Installing E modules to the patient monitor 1 Connect a module by al...

Страница 72: ...on if available All devices of a system must be connected to the same power supply circuit Devices which are not connected to the same circuit must be electrically isolated when operated CAUTION POWER REQUIREMENTS Before connecting the device to the power line check that the voltage and frequency ratings of the power line are the same as those indicated on the device s label If this is not the cas...

Страница 73: ...es that are specified compliant with IEC 60950 1 or IEC 60601 1 may be connected to the Ethernet MC or IX ports The patient monitor can be connected either to the wired MC Network or S 5 Network Connect the MC Network or the S 5 Network patch cable to the network connector labelled as MC in the rear panel of the patient monitor 6 7 Connecting to the IX Network Connect the IX Network patch cable to...

Страница 74: ...USB devices NOTE The USB connection to a D15K or D19KT secondary display reserves one of the USB connectors in the rear panel of the patient monitor Connect the following devices to the USB ports in the rear panel of the patient monitor or at the bottom of a D15K or D19KT display Alphanumeric keyboard Mouse Remote control Barcode reader NOTE Always use a GE supplied barcode reader for the CARESCAP...

Страница 75: ...le only if the anesthesia machine supports this feature NOTE The remote on standby function is disabled when the patient monitor is battery powered 6 12 Connecting a local printer to the IX connector You can connect a local laser printer directly to the patient monitor s IX connector with a crossover cable 1 Connect the Ethernet cable to the IX connector labelled as IX in the rear panel of the pat...

Страница 76: ...ltiple socket outlet supplying patient care equipment The use of multiple socket outlet for a system will result in an enclosure leakage current equal to the sum of all the individual earth leakage currents of the system if there is an interruption of the multiple socket outlet protective earth conductor WARNING Non medical equipment does not provide the same level of protection against electrical...

Страница 77: ...y needed only after replacing the uDOM or the touchscreen sensor Calibrate the connected touchscreens one by one as follows 1 Select Monitor Setup Service Calibrations 2 In the Enter Password dialog box type the username and password and press Enter Username biomed Password Change space Me NOTE Username and password are case sensitive NOTE The factory default password for the username biomed is Ch...

Страница 78: ...e patient monitor may be configured to the MC Network or IX Network or both NOTE The following IP addresses are reserved and not valid 172 16 254 254 172 16 255 255 172 18 254 254 172 18 255 255 192 168 250 x 192 168 252 x 192 168 253 x 192 168 254 1 and 192 168 254 2 5 In the Network Configuration window below MC Network a Enter a Static IP address b Enter a valid Netmask level c Select the appli...

Страница 79: ...nfiguration a Enter a Static IP address b Enter a valid Netmask level c Enter a valid Default Gateway d Enter valid DNS Server addresses if applicable 3 Select the applicable Speed Duplex option 4 Select Save The S 5 virtual ID change will be applied immediately All other network configurations will be saved and become active after the patient monitor is restarted 7 2 4 Configuring WLAN NOTE The n...

Страница 80: ...r multipath interference When set to Enabled Antenna Diversity monitors the signal from each antenna and automatically switches to the one with the better signal 5 Configure the appropriate WLAN settings Note that the configurable attributes and available values are different for MC Network and S 5 Network ...

Страница 81: ...shold is a numeric value within the range of 0 to 2347 NOTE This option is not configurable in the S 5 Wireless Network Fragmentation Threshold Configure the Fragmentation Threshold value Fragmentation Threshold specifies the maximum frame size a wireless device can transmit without fragmenting the frame A valid Fragmentation Threshold is a numeric value within the range of 256 to 2346 Use the def...

Страница 82: ...re the DSCP settings for the following network traffic Real Time Clinical Traffic Non Real time Clinical Traffic Non Real time Non Clinical Traffic Valid DSCP ranges Real Time Clinical Data 32 47 Waveforms Parameters etc Non Real time Clinical Data 0 7 Trends Full Disclosure Printing etc Non Real time Non Clinical Data 8 23 InSite Service traffic etc DSCP Differentiated Services Code Point setting...

Страница 83: ...infrastructure NOTE The authentication method used with WEP is Open System Authentication NOTE The wireless MC Network supports only confidentiality methods None WEP 128 bit WPA PSK TKIP and WPA2 PSK AES CCMP NOTE The S 5 Wireless Network supports only confidentiality methods None WEP 64 bit and WEP 128 bit Key Index Select the Key Index to be used The Key Index of the wireless client must match t...

Страница 84: ...work 1 Log in to Webmin 2 Select Configuration Time Pass Phrase Key Enter the Pass Phrase Key for the used confidentiality method as an ASCII or HEX string as selected above WEP 64 bit WEP 128 bit Enter a 5 13 character long ASCII string The system will generate a 10 26 character hexadecimal security key and assign it to the currently selected Key Index OR Enter a 10 26 character long hexadecimal ...

Страница 85: ...k and space The unit name may be up to seven characters long and bed name up to five characters long 4 Select Save The unit and bed name configuration takes effect immediately 7 5 Configuring printers You can configure the patient monitor to print to up to 12 laser printers connected on the IX Network Use the Webmin sub modules to install or delete a printer and to print a test page NOTE Ensure th...

Страница 86: ...s assigned to the printer to be deleted redirect the printout to another valid printer 1 Log in to Webmin 2 Select Configuration Printers 3 In the Sub Modules for Printers menu select Delete Printer 4 In the Delete Printer window select the printer to delete 5 Select Save The change will take effect immediately 7 5 3 Printing a test page 1 Log in to Webmin 2 Select Configuration Printers 3 In the ...

Страница 87: ...defined prefix that is used as the first two characters in a temporary patient ID to ensure its uniqueness inside the hospital 1 Log in to Webmin 2 Select Configuration Admit Settings 3 In the Sub Modules for Admit Settings menu select Patient ID Prefix 4 In the Patient ID Prefix window enter a 2 character prefix NOTE Valid values are uppercase letters and numbers 5 Select Save All changes take ef...

Страница 88: ...ation 7 7 3 Configuring Length Delimited Parser information Points to note If you configure Age you must either select the Age Unit item or one of the age units e g Years Months Weeks Days below Fixed Option If you configure Height you must either select the Height Unit item or one of the height units e g Feet Inches Meters Centimeters Millimeters below Fixed Option If you configure Weight you mus...

Страница 89: ... information In this example the barcode contains 10 items The following table lists the starting position and length of each item The following sample shows the corresponding entries in the Admit Discharge Configuration window Item Item selection in the top part of the screen Fixed Option selection Height Unit Both Height and Height Unit Non Fixed Height only Select value from drop down list Weig...

Страница 90: ...ure Height you must either select the Height Unit item or one of the height units e g Feet Inches Meters Centimeters Millimeters below Fixed Option If you configure Weight you must either select the Weight Unit item or one of the weight units e g Kilograms Grams Micrograms Pounds Ounces below Fixed Option For an example of admit discharge configuration for a length delimited parser see Example of ...

Страница 91: ...ode it will be misinterpreted as a field delimiter 3 Below Gender Code enter the codes that identify Male and Female 4 Below Fixed Option select the applicable value 5 Scroll to the bottom of the window and select Save All changes take effect immediately Example of character delimited parser information entry In the following example the barcode contains 12 items and uses the pound sign as a delim...

Страница 92: ...ple shows the corresponding entries in the Admit Discharge Configuration window 7 7 5 Barcode data specifications Points to note The maximum length of the entire barcode is 300 Height 8 Height Unit 9 Weight 10 Item Sequence number of the item in the barcode ...

Страница 93: ... of Birth and Year of Birth also need to be present Year of Birth 4 1880 to current year If Year of Birth is present in the barcode then Day of Birth and Month of Birth also need to be present Age 99 Numeric 9999 9999 Either a period or comma is accepted as a decimal symbol Age Unit 99 A Y YR YRS years MO MOS months WK WKS weeks D DAY DYS days Gender 1 If gender is configured not fixed this must b...

Страница 94: ...ent monitor is restarted The keyboard locale takes effect immediately after it is submitted 7 10 Selecting national requirements Activate France specific defaults for the ECG HR adjustment range and the reminder beep behavior 1 Log in to Webmin 2 Select Configuration National Requirements 3 In the National Requirement window select the applicable option 4 Select Save The national requirements chan...

Страница 95: ... be changed To set the host asset number 1 Log in to Webmin 2 Select Configuration Host Asset Settings 3 In the Host Asset Settings window enter the user assigned host asset number up to 32 ASCII characters long in the Change Value to column 4 Select Save The host asset changes take effect immediately 7 13 Changing passwords You can change the passwords for the biomed and clinical users Webmin sub...

Страница 96: ... immediately 7 14 Restarting the patient monitor You can use the Restart module in Webmin to restart the patient monitor after making configuration changes that require a manual restart to come into effect For example after changing network or language settings or adding activation codes for licenses NOTE Loss of monitoring This function is enabled only when the patient monitor is in a discharged ...

Страница 97: ... for the Remote Service communication to work Otherwise select None If the HTTP proxy server requires user authorization a specific Username and Password is required Otherwise select None These values are determined by the customer Port Username and Password Remote Service Configuration Item Description Comments System ID Identifies the system to the GE back office servers These values are read on...

Страница 98: ...nsfer clinical settings from a patient monitor with OR and PACU software only to other patient monitors with OR and or PACU software not to patient monitors with ICU NICU and ED software 7 16 1 Saving settings NOTE The patient monitor must not be configured to receive any alarm notifications from other monitors when the settings are saved Before saving the settings to a patient monitor that is con...

Страница 99: ...ame or select Browse to select a file from the Choose File to Upload dialog box Select Upload to load the settings b If you are using Webmin locally through the integrated browser you can load the settings file from a USB flash drive that is connected to one of the patient monitor s USB ports The Load Settings window will show you the available settings files Select the settings file to be loaded ...

Страница 100: ...t patient monitor NOTE Check and manually modify the following WLAN settings after activating them in the destination monitor RTS Threshold Fragmentation Threshold TXOP for Best Effort TXOP for Background Select Factory default settings to restore all platform and clinical settings back to the factory defaults Select US default clinical settings to restore all clinical and platform settings back t...

Страница 101: ...activation is completed and the patient monitor has performed an automatic restart d Verify that the settings activation was successful and the patient monitor is running the activated settings 7 16 4 Canceling pending settings activation You can cancel pending settings activation anytime while the message Setting activation after next case end Setting activation after next discharge is shown on t...

Страница 102: ...ackage select the desired radio button in the Active column 7 Select Save All license changes take effect after the next patient monitor restart NOTE The software license for the active host software version is shown in the Software Package window below the Active Software License This license is typically entered before activating a new host software version see Software Management WARNING If the...

Страница 103: ...oftware into the inactive memory partition of the patient monitor Transfer software using the GE Healthcare Software Transfer Utility that runs on a service PC With this application you can transfer new software from a software CD to the patient monitors over the CARESCAPE Network IX IX Network or a crossover cable The new transferred software is inactive in the patient monitor s until you activat...

Страница 104: ... more information refer to the latest version of the CARESCAPE Modular Monitors Supplemental Information Manual Contact GE Healthcare for any inquiries regarding the software CD and or activation code for the new host software version 1 Log in to Webmin 2 Select Configuration Software Management The Software Management window displays 3 In the Software List select the software that you want to act...

Страница 105: ...y 4 In Erase Inactive Software After Activation select 5 Select Activate to start the host software activation 6 Confirm activation by selecting Yes in the Host Software window that opens The patient monitor shows a screen saver that informs about the ongoing software activation No to keep the currently active software version as inactive software after the new software version is activated succes...

Страница 106: ...monitor restarts automatically 8 Verify that the software activation is successful and the patient monitor runs the activated software Activating host software after next case end discharge If you want to activate the new host software with a time delay that is after the next case end patient discharge follow this instruction You can initiate the software activation process with a delay during an ...

Страница 107: ...complete and the patient monitor restarts automatically If the patient monitor starts up normally and no error messages appear on the display the activation is successful Activating module software Follow this instruction when you want to activate new sGAS software to a CARESCAPE Respiratory Module activate new PDM software to a CARESCAPE Patient Data Module activate new PiCCO software to a E PiCC...

Страница 108: ... display 1 Log in to Webmin 2 Select Configuration Software Management The Software Management window opens 3 Select Host Software from the Software List and select Next 4 In the Host Software window below Status verify that the Software activation pending is Yes 5 Select Cancel The pending software activation is cancelled and the patient monitor continues running the current software version 7 18...

Страница 109: ...ion to the installed monitoring system Carefully inspect the patient monitor and the connected peripheral devices for any damage Verify that the patient monitor and the connected peripheral devices are properly mounted with specified mounting solutions Verify that the cables between the patient monitor and the connected peripheral devices are intact properly connected and secured to the right conn...

Страница 110: ...ate unless a properly wired power outlet is used 8 2 3 Power cord and plug Verify that the power cord being used with the patient monitor is good To do this check the following Inspect the power cord for wear or damage regularly If damage is suspected test for continuity through each conductor of the power cord connector Replace the power cord as necessary with a regulatory approved cord for the c...

Страница 111: ...urrent source with a frequency of 50 or 60 Hz with a no load voltage not exceeding 6 V is passed for at least 5 seconds but not more than 10 seconds through the protective earth terminal or the protective earth pin in the mains plug and each accessible metal part which could become live in case of failure in basic insulation 2 The voltage drop between the parts described is measured and the impeda...

Страница 112: ...his test NOTE The measuring device MD represents the network and voltage measuring instrument and its frequency characters according to IEC 60601 1 1 Configure the safety analyzer as follows NC Polarity NORMAL Neutral CLOSED 2 Power on the device under test 3 Read and record the current leakage indicated on the safety tester 4 Configure the safety analyzer as follows SFC Polarity NORMAL Neutral OP...

Страница 113: ...ngs shall be less than or equal to 1 mA 8 2 6 Enclosure leakage current touch current test This test measures current leakage through the exposed conductive parts on the device under test Perform the test in Normal Condition NC and in two different Single Fault Conditions SFC 1 earth open and 2 one of the supply conductors open at a time Perform the test with normal and reverse polarity NOTE Refer...

Страница 114: ...arity REVERSED Neutral CLOSED Earth GND OPEN 9 Read and record the current leakage indicated on the safety tester 10 Configure the safety analyzer as follows SFC Polarity REVERSED Neutral OPEN Earth GND CLOSED 11 Read and record the current leakage indicated on the safety tester 12 Configure the safety analyzer as follows NC Polarity REVERSED Neutral CLOSED Earth GND CLOSED 13 Read and record the ...

Страница 115: ... patient connector refer to the service instructions included in the safety test body kit NOTE If not otherwise stated in the table below each test body is connected directly to the specified connector in the patient module Table 1 Patient connectors to be tested with each module Module Patient connector E PRESTN E PSMP E PSMPW ECG SpO2 E RESTN E PSM E PSMW ECG SpO2 E PRETN ECG P1 E P E PT P3 P7 E...

Страница 116: ...g instrument and its frequency characters according to IEC 60601 1 NOTE Perform this test for all the connected parameter modules and patient connectors specified in Table 1 1 Configure the safety analyzer as follows NC Polarity NORMAL Neutral CLOSED Earth GND CLOSED 2 Power on the device under test 3 Read and record the current leakage indicated on the safety tester 4 Configure the safety analyze...

Страница 117: ...14 Power off the device under test 15 Repeat this test for all the connected parameter modules and patient connectors specified in Table 1 Acceptance criteria in Normal Condition NC All readings shall be less than or equal to 10 μA Acceptance criteria in Single Fault Condition SFC earth open or one of the supply conductors open at a time All readings shall be less than or equal to 50 μA 8 2 9 Pati...

Страница 118: ...current leakage indicated on the safety tester 6 Power off the device under test 7 Repeat this test for all the connected parameter modules and patient connectors specified in Table 1 Acceptance criteria All readings shall be less than or equal to 50 μA 8 2 10 Test completion 1 Disconnect the safety analyzer from the power outlet 2 Disconnect the test equipment from the patient monitor 3 Disconnec...

Страница 119: ...d all images are clear 2 Verify that the brightness is good Adjust if necessary Touchscreen control Perform this test only if the integrated main display or the optional secondary display is a touchscreen 1 Verify the operation and the calibration of a touchscreen by touching a corner of an active parameter window Verify that the selected menu is opened 2 Select to close any open menu and return t...

Страница 120: ...vity is started 3 Select to close any open menu and return to the main display 8 3 6 Mouse Perform this test only if a mouse is connected to the patient monitor 1 Move the mouse until the pointer arrow is over an active parameter window or a main menu item you wish to select and click the left mouse button once to select it 2 Select to close the open menu and return to the main display 8 3 7 Alpha...

Страница 121: ... is connected to a wireless MC Network or S 5 Network NOTE The wireless network must be properly installed and the patient monitor must be within the wireless coverage area Parser Type Test Procedure Length Delimited or Character Delimited Parser 1 2 3 4 Select Scan from Barcode 5 Scan a known test barcode obtained from the hospital NOTE The barcode data content must be known and in compliance wit...

Страница 122: ... the MC Network cable back to the MC Network connector if applicable 8 3 11 IX printers Perform the following test only if the patient monitor is connected to a printer in the IX Network and you did not print a test page while you configured the IX printer 1 Log in to Webmin 2 Select Configuration Printers Print Test Page 3 Select the IX Printer to test 4 Select Save 5 Verify that the test page wa...

Страница 123: ...ƚ ďƌĞĂŬĞƌ ĞWŽƌƚ ƌĞƐŝƐƚŽƌ ĚĞĐŽĚĞ W D ŽĐŬŝŶŐ ƉĞĂŬĞƌ hŶŝƚ K ƚĂŶĚĂƌĚ K ĚǀĂŶĐĞĚ K h Ϯ ϰ s Ͳ ϭ ϭ ϭϬ ϭϬϬ Ͳd ϭ ϰ ϵͲĞWŽƌƚ Ϭ ϭ ZĞŵŽƚĞͲKE Ϭ ϭ t E dƌĂŶĐĞŝǀĞƌ ĂƌĚ zW ϯ Žƌ zW ϭϰ DĞŵďƌĂŶĞ ŬĞLJƉĂĚ ŝŶŐĞ ůĞdž ŽĂƌĚ ƚŚĞƌŶĞƚ DŽĚƵůĞ ƵƐ ŽŶƚƌŽůůĞƌ D Ϳ DŽĚƵůĞ ƵƐ ŶƚĞƌĨĂĐĞ ZĞĐŽƌĚĞƌ ŝŶƚĞƌĨĂĐĞ ĞĨŝď Θ W LJŶĐ ŶƚĞƌĨĂĐĞ DĂŝŶ WƌŽĐĞƐƐŽƌ DĞŵŽƌLJ ůĂƐŚ Z DͿ tĂƚĐŚĚŽŐ Θ sŽůƚĂŐĞ ƐƵƉĞƌǀŝƐŽƌ ZĞĂů dŝŵĞ ůŽĐŬ sŝĚĞŽ ŽŶƚƌŽůůĞƌ dŽŶĞ ĞŶĞƌĂƚŽƌ Θ ƵƉĞƌǀ...

Страница 124: ...0 mAh lithium ion battery located in the battery compartment The battery cable connects the battery board to the DC DC board The battery pack supports Smart Battery Data and it communicates with the DC DC board using the System Management Bus SM Bus Battery management is handled by the DC DC board 9 8 9 8 RDUG EORFN GLDJUDP 9 DF GF XVH 9 9B30 15 2 5 6 5 200 21 67 75 5 027 B21 6 6B 855 17 5V 02 B 8...

Страница 125: ...nagement Controller controls power supplies sequencing It measures board temperature and voltages by an internal A D converter System and module currents also are measured and the corresponding power consumptions calculated PMC communicates with the CPU board via serial communication and with the battery and battery charger IC via SM Bus I2C bus Inter Integrated Circuit is used for communication w...

Страница 126: ...FORFNV 2 6HULDO FRQQHFWRU 86 86 DXGLR FRGHF 3 0 9 ODVV XGLR PSOLILHU 0 3 ILOWHU 6SHDNHU RXW 86 21 67 B 87721 2 2 1B 7 9 9 9 9 9 RQQHFWRU WR 0RGXOH IUDPH ERDUG HYHORSPHQW FRQQHFWRU 7 3RUW 23 7 VLJQDOV 3 5HJ QW 96 1 WR LQYHUWHU QDORJ 9 9 9 9 DWFKGRJ 7 5HVHW 60 ZLUH VHULDO SRUW 606 5 B7 606 5 B5 3RZHU PDQDJHPHQW 977 5 B72B30 7 B 520B30 9 9 7 5 71 66 57 5 9 XIIHU 3 B5 3 9 9 96 6 5 027 B21 1B 7 21 67 B...

Страница 127: ...nvasive pressure outputs and a digital marker out signal for a user supplied equipment e g a defibrillator or an intra aortic balloon pump Supply voltages The CPU has converters to generate 1 5 V 1 8 V and 2 5 V supply voltages for the processors and 12 V and 12 V voltages for the sync connector signals from the 3 3 V received from the DC DC board The CPU board passes the supply voltages created b...

Страница 128: ...t unit and a Backlight LED Adapter Board The earlier production series have an LCD display with 2 CCFL Cold Cathode Fluorescent Lamp backlight units and a Backlight Inverter Board To identify the display type of the patient monitor detach the top cover and check the color of the backlight cable and the rear of the LCD panel LCD display with a LED backlight unit and a Backlight LED adapter board Th...

Страница 129: ... board has a micro controller that manages the following main functions It reads the user input from the keypad matrix and Trim Knob encoder and passes the information to the CPU board A separate touchscreen controller digitizes the user input received from the touchscreen sensor and passes the coordinate information directly to the CPU board It turns the alarm light LEDs on and off according to t...

Страница 130: ...WRU 3 0 32 5B 85 B9 68 B 50 0 B 8 ERDUG EORFN GLDJUDP YVG hƐĞƌ ŶƚĞƌĨĂĐĞ ŽĂƌĚ ŵŝƚŚͲdƌŝŐŐĞƌ ďƵĨĨĞƌƐ yϭϬϬ DϮ Wh ŽĂƌĚ ŝŶƚĞƌĨĂĐĞ ĐŽŶŶĞĐƚŽƌ ĚƌŝǀĞƌ ĐŝƌĐƵŝƚƐ dϯϱϵϭ h ƉŽƌƚ нϱs нϯ ϯs ZĞƐĞƚ ͺ dͺ ͺWd Kͺ ͺ ͺWd KE d zͺ hddKEͺWd ͺWKtͺ ͺWd ŝůƚĞƌƐ ĞLJďŽĂƌĚ DĂƚƌŝdž yͲůŝŶĞƐ ϰͿ ĞLJďŽĂƌĚ DĂƚƌŝdž zͲůŝŶĞƐ ϰͿ K WtDϯ нϯ ϯs dƌŝŵŬŶŽď ĚŝƌĞĐƚŝŽŶ ƐŝŐŶĂů нϯ ϯs ZYϬ ZYϭ K d EK ͺ d EK ͺ d EK ͺW dƌŝŵŬŶŽď ĚŝƌĞĐƚŝŽŶ ƐŝŐŶĂů dƌŝŵŬŶŽď ƉƵƐŚď...

Страница 131: ...w and blue LEDs to display different priority visual alarms The left hand side of the alarm light board has blue LEDs for the silence alarm indicator light The alarm light board also has a separate light sensor that is capable of measuring the ambient light intensity The ambient light feedback can be used to auto adjust the brightness of the keypad backlits alarm light LEDs and the LCD display bac...

Страница 132: ...s integrated to the microprocessor in the CPU board The Advanced Interface Board passes the remote on signal i e an on standby signal from an external device to the PMC CPU in the DC DC board The operation of the remote on connector is disabled when the patient monitor is battery powered The Advanced Interface Board has I2C A D converter and I2C temperature sensor to measure the on board VSYS and ...

Страница 133: ...ot connected to a wired network through the MC port 9 1 6 Pivoting module frame Hinge flex board The hinge flex board interfaces the E module interface board and the module frame flex board to the CPU board 0 237 2 ERDUG 0 38 ERDUG FRQQHFWRU LOWHUHG 70 6 9 9 LOWHUHG DQDORJ 0 0 WR 0 7UDQVPLW 5HFHLYH FORFNV 86 IURP 0 38 9 9 LOWHUHG 86 0 P FXUUHQW OLPLW 9 XDO 86 LOWHULQJ DQG 6 SURWHFWLRQ 9 9 LOWHULQJ...

Страница 134: ...ePort and recorder 5V REC and 12 V REC supply voltages This information is communicated to the PMC CPU in the DC DC board over the I2C bus E module interface board The optional E module interface board connects to the hinge flex board It has two D25 male connectors that provide an interface for two single width E modules or for one double width E module The E module interface board provides the fo...

Страница 135: ... supply voltage for the recorder from the 12 V received from the DC DC board PDM interface The detachable PDM docking station interfaces to the module frame flex board The 10 pin PDM connector provides the VSYS ePort supply voltage and the Ethernet communication lines to the PDM module PSM interface The PSM connector is part of the module frame flex board The PSM connector provides the 15 VMod PSM...

Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...

Страница 137: ...odule Frames and Modules Technical Manual for the battery maintenance procedure of the PDM battery Corrective maintenance Service personnel shall perform the following checkout procedure steps after any corrective maintenance before taking the monitor back into clinical use Performed service activity Required checkout procedure Visual inspections section 10 1 Electrical safety test section 10 2 Fu...

Страница 138: ...Knob Encoder FRU Keypad FRU Yes Yes 10 3 1 Start up 10 3 5 Keypad and remote After detaching or replacing LCD Display Unit FRU Backlight LED Adapter Board FRU or Backlight Inverter Board FRU Yes Yes 10 3 1 Start up 10 3 2 Display 10 3 3 PSM PDM identification After detaching or replacing Rear Unit Assembly FRU AC DC Power Supply Unit FRU DC DC board FRU CPU timekeeper battery FRU Hinge Flex Board ...

Страница 139: ...re the ECG1 waveform field and the NIBP parameter window to the patient monitor screen with adequate priority Connect a PSM or PDM module to the patient monitor 2 Verify that the ECG waveform field the NIBP parameter window and the related information appear on the patient monitor screen 10 3 4 E module identification 1 Log in to Webmin 2 Select Information Device Information 3 Verify that the inf...

Страница 140: ...veform 2 V1 3 Select Monitor Setup Printing Waveforms Print Waveforms or the related soft key in the main menu or the Print Waveforms hard key in the keypad to start printing 4 Verify that the recorder starts printing Let the recorder print for approximately 10 seconds and verify the following things from the printout The header line contains the date time and some other applicable status and conf...

Страница 141: ...G trunk cable to the green ECG connector in the module 3 Connect the 5 leadwire sets to the trunk cable and to the simulator 4 Connect the invasive pressure adapter cable to the simulator and the other end to the module 5 Connect the analog output cable to the defibrillation synchronization connector in the patient monitor 6 Connect the open wires of the other end of the analog output cable to the...

Страница 142: ...figure invasive pressure Sensitivity 5 µV V mmHg InvBP output 0 mmHg static pressure or atmosphere Testing the synchronization connector 1 Test the ECG analog output signal a Connect the oscilloscope probe to the ECG analog output signal grey wire and analog ground blue or white wire b Adjust the oscilloscope Time and Volts scales c Verify that the ECG analog output waveform shown on the oscillosc...

Страница 143: ...he oscilloscope probe to the marker out signal black wire and common ground red wire b Change the oscilloscope Volts scale c Verify that the signal shown on the oscilloscope screen corresponds to the picture above 4 Test the pulse width of the Marker out signal a Change the Time scale in the oscilloscope b Verify that the signal shown on oscilloscope screen corresponds to the picture above IP anal...

Страница 144: ...y is new Full charge capacity The actual amount of charge the battery can store and deliver Remaining charge capacity The amount of full charge capacity currently remaining in the battery This is a percent of full charge capacity 10 4 1 Use recommendations GE recommends the following methods to improve battery performance Location Position the equipment in a location that does not artificially inc...

Страница 145: ... outlet and turn on the patient monitor Leave the patient monitor on until the battery is fully recharged and the orange battery charging indicator LED turns off 3 Repeat the steps 1 and 2 once The battery is now conditioned and ready for use However if the Condition monitor battery message is still shown on the patient monitor screen repeat the conditioning cycle once more If the problem persists...

Страница 146: ...countries USA You may follow the battery manufacturers instructions on the battery to recycle it Alternatively you may return GE product batteries to GE for recycling For information about returning batteries to GE contact your authorized GE Service representative or contact GE Equipment Services at 1 800 437 1171 Canada Contact the approved battery stewardship program in your province for informa...

Страница 147: ...leshooting complete a thorough visual inspection to be sure that There is no physical damage All peripheral devices are connected properly The patient monitor and the connected peripheral devices are properly powered Also verify that the problem is not caused because of 1 Incompatibility issue See the patient monitor s supplemental information manual for the list of compatible devices 2 Incorrect ...

Страница 148: ...formation and device information 11 2 1 Configuration information The Configuration Information module shows the current platform configuration of the patient monitor and the connected peripheral devices To view configuration information 1 Log in to the Webmin 2 Select the Information tab 3 Select Configuration Information ...

Страница 149: ...tus enabled disabled or trial feature code and the expiration date for a trial license Default Clinical Settings Current default clinical settings Admit Settings Patient ID Prefix Unit and Bed Name Unit name and Bed name for CARESCAPE Network E Module Information STP TP ST configuration information and P PT PP configuration information from E modules S 5 Printers Printer name IX Printers Printer n...

Страница 150: ...on method Confidentiality method and Key Index Device information Host Information Active software part number and version Inactive software part number and version Host serial number Host asset number MC Network IP address IX Network IP address MAC address S 5 Network virtual ID CPU hardware version PMC hardware version UIC hardware version PMC software version UIC software version and EMBC softw...

Страница 151: ... 5 Printers Printer name Installed IX Printers Printer name hostname or IP address Printer Location Information Printout type Alarm Waveforms Numeric Trends Reports and Waveforms and Printer location PDM License Information PDM license option status and number of licenses UNITY ID Information Product ID Unity Network ID software number and version Date Time Device name and software version of each...

Страница 152: ...ed from the CPU board AC DC voltage mV The AC DC voltage is generated by the AC DC power supply unit and supplied as an input voltage to the DC DC board The voltage is measured from the DC DC board The AC DC voltage is 0 if the patient monitor is not connected to the AC mains Also see section 11 4 Power management LEDs VSYS voltage mV 15V MOD voltage mV 12V voltage mV 5V voltage mV 3 3V voltage mV...

Страница 153: ...age is enabled The VSYS RACK voltage is measured from the module frame flex board before the VSYS ePort circuit breaker The VSYS EMOD voltage is measured from the E module interface board before the VSYS EMOD circuit breaker The VSYS EMOD voltage is shown as if the E module interface board is not installed The VSYS OPTION voltage is supplied for a PDM module that is connected to the ePort DB9 inte...

Страница 154: ...ormation about the WLAN driver WLAN status and the detected access points 1 Log in to Webmin 2 Select Diagnostics WLAN 3 Scroll down the page to view the following WLAN diagnostics information NOTE Some WLAN data is not updated until the monitor connects to an access point Log Contents WLAN Driver Driver API Version Driver Version Firmware Version MAC Address ...

Страница 155: ...gory Voice Video Best Effort Background The DSCP settings for network traffic types Realtime Clinical Traffic Non Realtime Clinical Traffic Non Realtime Non Clinical Traffic SSID i e Service Set Identifier network name Security status Authentication method i e open Confidentiality method i e None WEP 64 WEP 128 WPA PSK TKIP WPA2 PSK AES CCMP Used Key Index if WEP 64 or WEP 128 is used RF readings ...

Страница 156: ... maintain application performance Operating channel number MAC Address NOTE If the patient monitor has associated with an Access Point information on only the corresponding Access Point will be displayed Log Contents Association Authentication state Description of the state Disconnected The WLAN client radio is not authenticated or associated to any network infrastructure Authenticating The networ...

Страница 157: ...tion occurred Module software activation information including the module type and serial number the type and version of the activated module software and the date and time the module software activation occurred Settings transfer history information including the type of the transferred settings the origin and destination and the date and time the settings transfer occurred Log file transfer hist...

Страница 158: ...s log file to GE Service for further investigation To view log files 1 Log in to Webmin 2 Select Diagnostics View Logfiles PDM log All PDM errors and messages System log OS events and errors including operating system related information such as clinical application startup and recovery information power on self test results etc UIC PMC WLAN and battery events and errors InSite with ExC events and...

Страница 159: ...ard are visible through the ventilation holes in the rear of the patient monitor see picture below LED Description On standby red The red on standby LED should be lit momentarily when the on standby button is pressed A lit LED normally indicates that the DC DC power supply board recognizes that the user has pressed the on standby button and should turn the patient monitor On Off If the On standby ...

Страница 160: ...AC mains even if the patient monitor is in standby mode A lit LED indicates that the DC DC board receives the 15 3 16 VDC from the AC DC power supply unit If the ACDC LED is not lit when the patient monitor is connected to live AC mains check the patient monitor for one of the problems Main fuses are blown DC DC cable is disconnected from DC DC board Faulty AC DC power supply unit especially if th...

Страница 161: ...tery is not connected to the patient monitor Slot status No battery Battery is not connected to the patient monitor No communication Battery is connected to the patient monitor but battery communication failure error condition is on Failure Battery error condition is on Discharging Patient monitor is operating on battery power Charging Patient monitor is connected to AC mains and charging the batt...

Страница 162: ...ery N A Battery is not connected to the patient monitor Field Description Remaining capacity mAh Remaining capacity of the battery in mAh Full capacity mAh Full capacity of the battery in mAh Full capacity compared to new Full capacity of the battery compared to the nominal full capacity of a new battery Cycle count The total count of charging and discharging cycles of the battery Voltage V Batter...

Страница 163: ...ot obstructed 4 Ensure that the patient monitor is installed to a location that meets the specified environmental requirements of operating temperature 5 Investigate the patient monitor thoroughly for potential short circuits and other electrical faults 6 If possible log in to Webmin and select Diagnostics Hardware Stats to identify the root cause for the error message If temperature is too low or...

Страница 164: ...f and then on again If the error situation reoccurs contact your local service representative Error 2 PMC update requires a mains supply Plug in the power cord and turn the monitor off and then on again PMC software update to the DC DC board has failed because the patient monitor was running on battery power The patient monitor shall be connected to the AC mains Plug in the power cord and turn the...

Страница 165: ...lect Audio Pause to acknowledge the message Replace monitor battery The full capacity of the battery is less than or equal to 50 compared to the full capacity of a new battery Replace the monitor battery with a new one Battery failure Battery failure Replace battery with a new one Condition monitor battery Battery needs conditioning Condition the monitor battery See section 10 4 5 Conditioning a b...

Страница 166: ...faulty Check the power cord for wear and damage and replace if necessary The power outlet does not meet specified requirements Check the power outlet being used Refer to the patient monitor s supplemental information manual for power requirements Check the power outlet being used see section 8 2 2 The cable between the AC DC power supply unit and the DC DC board is loose or faulty Check that the c...

Страница 167: ...tient monitor when it is both connected to AC mains and a fully charged monitor battery is installed The cable between user interface board and CPU board is loose or faulty Check the status of the red ON Stdby button LED on DC DC board when On Stdby button is pushed During the push LED should illuminate If it is not illuminating the cable between the user interface board and DC DC board is loose o...

Страница 168: ...r board cable is intact and properly connected to the DC DC board and the backlight LED adapter board Reconnect the inverter cable if loose Replace the inverter cable if faulty The backlight LED adapter board is faulty Replace the backlight LED adapter board The backlight cable between the LED backlight unit and the backlight LED adapter board is loose or faulty or The LED backlight unit is faulty...

Страница 169: ...conds 3 Re install the batteries and plug the power cord back to the monitor 4 Turn the monitor on and wait until the monitor restarts automatically Webmin login screen appears after the startup screen There is no host software in uDOM Transfer and activate host software to monitor see section 7 18 or replace the uDOM Contact GE service for support Problem Possible cause Recommended actions Contin...

Страница 170: ...s normal operation The alarm light illuminates during the power on self test only if it is a cold start more than 15 minutes from the previous start up Problem Possible cause Recommended action Wrong character is displayed when a key is pressed on keyboard The keyboard locale is not configured correctly Configure the keyboard locale correctly see 7 9 Wrong character is displayed when a barcode is ...

Страница 171: ...onnected to network CPU timekeeper battery empty Replace CPU timekeeper battery Time not configured properly Configure date and time see 7 3 System time is incorrect when patient monitor is connected to MC Network Network device time synchronization error When adding a new device to the CARESCAPE Network the existing devices on the CARESCAPE Network will synchronize to the new device s time To pre...

Страница 172: ...ption and select Activate Unable to upload a license file The license file is corrupted The license file is for a patient monitor with a different serial number Log in to Webmin Configuration Licenses If you have printed license information select Software Package and Host Licensing If you have a license file select Upload License The wrong software application is displayed on the patient monitor ...

Страница 173: ... most likely faulty Refer to Module Frames and Modules Technical Manual for troubleshooting instructions Problem Possible cause Recommended action Recorder does not work Graph location is not configured correctly Check the configuration Monitor Setup Printing Devices Setup Recorder board failure Check the status of the 5 V REC and 12 V REC voltages see section 11 3 1 Hardware statistics If one of ...

Страница 174: ...MBC section of the CPU board Check EMBC Frame logs for possible EMBC failures see section 11 3 4 If the cause is none of the above the problem is most likely in the CPU board Replace the CPU board Possible cause Recommended action Pole mount cable loose or faulty if the PSM module is mounted to a pole mount Check that the Module Bus Adapter for PSM and the connection cable is intact and properly c...

Страница 175: ...that the supply voltages are within specification VSYS voltage mV VSYS_EPORT mV VSYS_PDM mV If PDM is connected to PDM dock DC DC board should identify PDM and enable the VSYS_PDM circuit breaker on CPU carrier board to connect VSYS voltage to VSYS_PDM If PDM is connected to ePORT DC DC board should identify PDM and enable the VSYS_EPORT circuit breaker on CPU carrier board to connect VSYS voltage...

Страница 176: ...dcasts from other network devices Hinge flex board is loose or faulty PDM is connected to the PDM docking station Check that the hinge flex board is intact and properly mounted to the CPU board and to the module frame flex board Module frame flex board faulty PDM is connected to the PDM docking station Check the status of the VSYS RACK and VSYS ePort see section 11 3 1 Hardware statistics If the V...

Страница 177: ...Troubleshooting 169 2062973 002 ...

Страница 178: ...DQ RX VHOHFW DQRWKHU SDWLHQW PRQLWRU WR YLHZ IURP WKLV SDWLHQW PRQLWRU HV QR 1HWZRUN SUREOHP URDGFDVW WUDIILF LV JRLQJ EHWZHHQ WKH SDWLHQW PRQLWRUV EXW QRW WR WKH 3UR FHQWHUV 7KH 3UR FHQWHUV PD QRW EH RQ WKH VDPH QHWZRUN DV WKH SDWLHQW PRQLWRUV 3OHDVH YHULI FRQILJXUDWLRQ VHWWLQJV XVLQJ WKH LUHOHVV 1 1HWZRUN RQILJXUDWLRQ XLGH DQG FRQVXOWLQJ WKH KRVSLWDO V 7 GHSDUWPHQW RQILJXUDWLRQ LVVXH 6HOHFW WKH ...

Страница 179: ...N SUREOHP 3OHDVH YHULI FRQILJXUDWLRQ VHWWLQJV ZLWK WKH KRVSLWDO V 7 GHSDUWPHQW HV 8VHU UURU 3OXJ LQWR FRUUHFW MDFNV QR RQILJXUDWLRQ LVVXH RUUHFW 3 DGGUHVVLQJ LVVXHV HV 1R WUDIILF LV JHWWLQJ WKURXJK 3RWHQWLDO QHWZRUN FRQILJXUDWLRQ LVVXH V WKH 66 WLHG WR WKH FRUUHFW 9 1 3 FRQILJ V WKH 9 1 EULGJHG WR WKH 5 6 3 1HWZRUN V WKH 3UR FHQWHU LQ WKH FRUUHFW 5 6 3 1HWZRUN V WKHUH D ILUHZDOO RU FFHVV RQWURO LV...

Страница 180: ... all bedside monitors affected The problem is related to the iCentral or to the cabling between the iCentral and the hub switch Proceed to the iCentral troubleshooting Check the cabling between the iCentral and the hub switch Bedside monitor doesn t register to the Monitor Network only one patient monitor or some of the bedside monitors affected The problem is related to the bedside monitor or to ...

Страница 181: ...ess network infrastructure IEEE 802 11b SSID mismatch between patient monitor and access point Enter correct SSID see section 7 2 4 Security settings mismatch between patient monitor and access point Incorrect or incompatible confidentiality method Incorrect Pass Key Incorrect key index Contact the hospital IT department for correct security settings and then update the patient monitor configurati...

Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...

Страница 183: ...ectronic assemblies being serviced Discharge any static charge you may have built up before handling semiconductors or assemblies containing semiconductors A grounded antistatic wristband or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors Use properly grounded test equipment Use a static free work surface while handling or working on assembli...

Страница 184: ...tached Avoid overtightening the screws as this may damage the existing thread patterns However when using battery operated tool ensure that it is equipped with torque limiter and the torque is properly adjusted 12 1 3 Required tools WARNING Due to possible high voltage present use an insulated screwdriver at all times 12 1 4 Before disassembly WARNING PATIENT MONITORING INTERRUPTION Make sure a pa...

Страница 185: ...me 6 Detach the monitor from the mount if installed NOTE If you disconnect the monitor from the AC power source and remove the battery prior to turning the monitor off the monitor will start alarming about power loss situation Silence this alarm by pressing the Silence alarm button When re installing the battery make sure that the charge level indicator LEDs are facing upwards and push the battery...

Страница 186: ...ŝŶŐ ĚĞĐŽƌĂƚŝŽŶƐ ŶƚĞƌĨĂĐĞ ďŽĂƌĚ ƌŽŶƚ ƵŶŝƚ ĂƐƐĞŵďůLJ t E ĂĐŬůŝŐŚƚ ĂĚĂƉƚĞƌ ďŽĂƌĚ ďŽĂƌĚ DŽĚƵůĞ ĨƌĂŵĞ ĂƐƐĞŵďůLJ W D ĚŽĐŬŝŶŐ ŵĞĐŚĂŶŝƐŵ DŽĚƵůĞ ĨƌĂŵĞ ĐŽǀĞƌ ƵŶŝƚ ͲŵŽĚƵůĞ ŝŶƚĞƌĨĂĐĞ ďŽĂƌĚ Ƶ KD dƌŝŵ ŶŽď ĞŶĐŽĚĞƌ ƌŽŶƚ hŶŝƚ ƐƐĞŵďůLJ Zh ƵƐĞƐ dŽƉ ĐŽǀĞƌ DŽĚƵůĞ ĨƌĂŵĞ ĨƌŽŶƚ ĐŽǀĞƌ ĞLJƉĂĚ hƐĞƌ ŶƚĞƌĨĂĐĞ ďŽĂƌĚ 12 4 1 12 3 2 12 3 3 12 3 9 12 4 2 12 3 4 12 2 1 12 2 5 12 3 6 12 2 4 12 3 5 12 2 2 12 4 3 12 4 3 12 2 6 12 3 10 12 ...

Страница 187: ...e so that the Module Frame faces to the front 2 Release the two snaps on the top of the cover unit press strongly with your thumb on the middle of the Module Frame cover unit s front edge pull out the front cover a little with the other hand 3 Release the four snaps in the inner side of the E module slot by pressing them inwards with your finger Release first the topmost two snaps and then the low...

Страница 188: ...n the Module Frame to a 90 angle so that the Module Frame faces to the front 2 Release the two snaps on the top of the cover unit press strongly with your thumb on the middle of the Module Frame cover unit s front edge pull out the front cover a little with the other hand 3 Press with finger from the top of the PDM docking rail ...

Страница 189: ...embly 181 2062973 002 4 Shift the front cover slightly down and to the left by pulling the hole in the bottom plate to release the two snaps on left side and one on the right side of the front cover 5 Detach the cover unit ...

Страница 190: ...moved Torque 1 2 Nm NOTE Be careful not to damage the Hinge Flex Board on the right of the recorder 3 Pull the recorder carefully out of the Module Frame NOTE If you find it difficult to detach you also can detach the recorder by removing the Modules Frame cover unit see section 12 2 4 4 Release the snaps in the recorder EMC cover with a flat blade screwdriver while simultaneously pulling the bott...

Страница 191: ...dule Frame as a whole NOTE Ensure that the recorder unit is firmly installed to the recorder connector and both snaps in the EMC cover lock properly 1 Release the two latches on the inner left side of the Module Frame by first pressing the snaps and then sliding the latches towards you 2 Turn the PDM Docking Mechanism to a 30 angle to detach it from the Module frame flex board connector Reassemble...

Страница 192: ...t Disassemble first a 12 2 1 Detaching the module frame front cover b 12 2 3 Detaching the PDM docking mechanism if installed 1 Release the two snaps inside the holes of the frame cover using a flat blade screwdriver 2 Pull the cover backwards to detach it Reassemble in reverse order ...

Страница 193: ...th a flat blade screwdriver and detach the plate 2 Disconnect the Hinge Flex Board from the Module Frame connectors NOTE Handle the flex board with care when disconnected Do not twist it or bend it down 3 Remove the six T10 screws mounting the E module Interface Board Torque 0 6 Nm NOTE When reassembling do not use excessive force to avoid damaging the plastic threads 4 Detach the board from the M...

Страница 194: ...c 12 2 4 Detaching the Module Frame cover unit d 12 2 5 Detaching the E module Interface Board steps 1 and 2 only 1 Remove the T10 screw Torque 0 6 Nm 2 Release the snap behind the flex cable Lift the flap with a flat blade screwdriver 3 Pull the Module frame Assembly to detach it from the Module Frame wall NOTE When reassembling check that the connector is intact Reassemble in reverse order ...

Страница 195: ...mains fuses are situated on the back of the monitor below the mains connector 1 Lever the screwdriver against the cable holder to release the fuse holder and pull it out of the frame 2 Pull the fuses out of the fuse holder and insert new ones 3 Push the fuse holder all the way in until it snaps to its place ...

Страница 196: ...sing the snaps with a flat blade screwdriver 2 Remove the six T10 screws that hold the Interface Board to the main unit Torque 0 8 Nm NOTE At first reassemble the screws in the middle 3 Pull the Interface Board out of the main unit with pliers NOTE Pull with the pliers from the plate on the interface board panel Reassemble in reverse order ...

Страница 197: ...t the bottom part of the decoration with a flat blade screwdriver Lift the decoration a little more to release the rest of the snaps and detach the decoration 2 To detach the Left Frame Housing Decoration Release the snap at the bottom part of the decoration by pressing it down with a flat blade screwdriver Lift the decoration a little more to unlock the rest of the snaps and detach the decoration...

Страница 198: ... the Module Frame to a 90 angle so that the Module Frame slot faces to the front 2 Remove the T10 screw located behind the Module Frame Torque 0 8 Nm 3 Detach the Top Cover by releasing the four snaps with a flat blade screwdriver Reassemble in reverse order Grease the rubber gasket of Top Cover to reassemble it smoothly ...

Страница 199: ... mechanism Consider detaching it according to section 12 2 3 2 Set the rear side of the Main Unit facing you 3 Remove the three T10 screws from the left side Torque 0 8 Nm 4 Remove the two T10 screws from the right side next to the Module Frame hinges Torque 0 8 Nm 5 Detach the Module Frame back plate and disconnect the Hinge Flex Board from the Module Frame connectors see section 12 2 5 NOTE Hand...

Страница 200: ... can detach the DC DC cable from the DC DC Board NOTE When reassembling check that the cable is connected properly 8 Detach the rear unit from the Front Unit Reassemble in reverse order NOTE The Front Unit is unstable when detached from the Rear Unit and may easily tilt forward Shock hazard Dangerous voltage ...

Страница 201: ...PU connector NOTE uDOM disk on module is the permanent memory of the patient monitor The software licenses and clinical and service configurations are stored to the uDOM Reassemble in reverse order uDOM replacement Contact GE Service if you doubt that the uDOM is defective and needs to be replaced Some of the configuration steps required after uDOM replacement can be performed only by GE service P...

Страница 202: ...over d 12 3 5 Disassembling the main unit into Rear and Front units 1 Detach the CPU timekeeper battery from the CPU board with a flat blade screwdriver Reassemble in reverse order To avoid network device time synchronization issues remember to readjust the time and date before you connect the monitor to network For details see 7 3 Setting time and date NOTE Dispose of the battery according to loc...

Страница 203: ...r 4 Detach the DC DC Board When reassembling Assemble the CPU Board first if detached Connect the Battery cable Slide the DC DC board to connect the CPU connector NOTE Do not bend the DC DC board when reassembling Connect the Inverter cable and the three T10 screws Torque 0 8 Nm Reassemble the Rear Unit and the Front Unit in reverse order NOTE After replacing the DC DC Board during the first start...

Страница 204: ...amage the Hinge Flex Board 4 Detach the Sync cable the Loudspeaker cable and the U I cable from the CPU Board 5 Remove the nine T10 screws that hold the CPU Board to the mid frame 6 Detach the CPU Board Torque 0 8 Nm Reassemble in reverse order Insert the WLAN card into the card slot Place the antenna cables carefully inside the cover NOTE Center align the user interface cable carefully to the con...

Страница 205: ...isassembling the main unit into Rear and Front units 1 Detach the AC DC cable from the cable holder 2 Detach the power cable from the AC DC Board 3 Remove the T10 screw that attaches the housing back cover to the Rear Unit Assembly Torque 0 8 Nm 4 Release the two snaps that are located on the left side next to AC DC board 5 Detach the housing back cover NOTE The loose cable clamp and spring may fa...

Страница 206: ...etaching the LCD Display Unit 6 Remove the four T10 screws mounting the AC DC board to the back plate Torque 1 6 Nm Hold the AC DC board to prevent it from falling Reassemble in reverse order Attach the 4 screws that mount the AC DC board Torque 1 6 Nm Ensure the cable clamp and the spring are in place Reattach the housing back cover so that the 4 snaps locks Attach the 1 screw that mounts the hou...

Страница 207: ...from the DC DC Board the Battery cable 3 Remove the five T10 screws with washer insulators mounting the Base Unit to the mid frame Torque 0 8 Nm 4 Remove the two T10 screws with washer insulators mounting the Base Unit to the mid frame Torque 0 8 Nm 5 Detach the mid frame from the Base Unit by lifting it up carefully CPU DCDC ...

Страница 208: ...ng Check that they go through the right holes in the mid frame when attaching the Base Unit Be especially careful to check the loudspeaker cable see the figure 6 To replace the mid frame detach DC DC CPU 12 4 3 Backlight LED Adapter Board Reassemble in reverse order DC DC Backlight CPU Board Board LED adapter board ...

Страница 209: ... 002 12 4 Front Unit Assembly 12 4 1 Detaching the Front Unit Assembly Disassemble first a 12 3 3 Detaching the Frame Side Housing Decorations b 12 3 4 Detaching the Top Cover 1 Remove the T10 screw at the bottom of the frame Torque 0 6 Nm ...

Страница 210: ... 5 Detach the Front Unit Assembly 6 Disconnect the U I cable from the User Interface Board Reassemble in reverse order NOTE Do not leave any dust or dirt between the LCD Display and the plastic display cover or the picture quality might deteriorate NOTE Center align the user interface cable carefully to the connector in the CPU board and or User Interface Board NOTE The keypad may be damaged when ...

Страница 211: ...c 12 4 1 Detaching the Front Unit Assembly 1 Remove the four T10 screws with washers two on both sides of the LCD Display and detach the display from the frame Torque 0 6 Nm 2 Detach the LVDS cable from the LCD display 3 Detach the Backlight Cable from the Backlight LED Adapter Board NOTE To get access to detach the cables you may have to lift the LCD display as showed in the pictures ...

Страница 212: ...or country laws 1 Remove the two screws T10 and PCB mounting parts holding the Backlight LED Adapter Board to the mid frame and detach the board 2 Detach the Inverter Cable from the Backlight LED Adapter Board Reassemble in reverse order NOTE Note the alignment pin to set the Backlight LED adapter board in the right place NOTE The Backlight LED Adapter Board FRU includes the inverter cable Prefera...

Страница 213: ...I cable Preferably use the original UI cable if it seems to be visually undamaged In case you need to replace the UI cable you need to additionally detach the rear unit assembly to get access to the CPU board and be able to connect the inverter cable to the CPU board NOTE After replacing the User Interface Board during the first start up the patient monitor will automatically check the UIC softwar...

Страница 214: ...the release liner of the keypad Place the hole at right end of the keypad to the Trim Knob shaft Then carefully guide the lower edge against the cover edging 3 Tear the membrane keypad from the Front Unit and detach the keypad connector that connects it to the User Interface Board Assembling Clean the glue from the surface Remove the release liner from the membrane keypad Connect the membrane keyp...

Страница 215: ...N is installed Detach carefully the WLAN antennas from the Front Unit NOTE Be careful not to bend the antenna boards or damage the antenna cables Reassemble in reverse order NOTE When replacing the Front Unit Assembly always replace the Keypad panel and keypad with new ones as they may damage when detached NOTE The antennas are attached to the Front Unit with double sided tape New tape is included...

Страница 216: ...ont Unit Assembly as shown in the picture Make sure the antenna board is firmly attached to the Front Unit Assembly Align the loop in the antenna cable carefully between the display and the aluminum Mid Frame as shown in the picture The orientation of the antenna cable is important to ensure reliable wireless communication 4 Make sure the contact area for the antenna board adhesive in the Front Un...

Страница 217: ...Disassembly and reassembly 209 2062973 002 NOTE Place the antenna cables carefully as shown in the pictures Reassemble the Top Cover Grease the rubber gasket of Top Cover to reassemble it smoothly ...

Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...

Страница 219: ... back plate with its mounting screw 5 M1133995 FRU Module Frame Assembly B650 Includes module frame assembly with module frame flex board and grounding plate 6 M1168210 FRU Standard Interface Board B650 Includes standard interface board 2 USB 1 RJ 45 1 DVI I related cover plate with labelling and 6 mounting screws 7 M1168215 FRU Advanced Interface Board B650 Includes advanced interface board 4 USB...

Страница 220: ...der B650 Includes the grey trim knob trim knob encoder w cable and mounting parts nut and washer 15a M1168227 FRU LCD Display Unit B650 Replaced with 2073132 001 15b 2073132 001 FRU LCD Display Unit Kit B650 01 Includes LCD display with LED backlight unit and 4 mounting screws Backlight LED adapter board and 2 mounting screws with PCB fastening Backlight cable Inverter cable Display LVDS cable 16a...

Страница 221: ...play and the 7 mounting screws and insulator plates to mount it to the bottom unit 27 M1168356 FRU Battery B650 Model Flex 3S3P rechargeable lithium ion battery 28a M1168339 FRU Cable Kit B650 for patient monitors that have a LCD display with CCFL backlights Includes the following parts Hinge flex board with ferrites Display LVDS cable with ferrites for displays with CCFL backlights User interface...

Страница 222: ...es the membrane keypad and the language specific keypad panel 14 keys 37 M1178181 FRU Keypad for Non Touchscreen model NL B650 Includes the membrane keypad and the language specific keypad panel 14 keys 38 M1178182 FRU Keypad for Non Touchscreen model DA B650 Includes the membrane keypad and the language specific keypad panel 14 keys 39 M1178183 FRU Keypad for Non Touchscreen model FI B650 Include...

Страница 223: ...model RU B650 Includes the membrane keypad and the language specific keypad panel 14 keys 48 M1178200 FRU Keypad for Non Touchscreen model JA B650 Includes the membrane keypad and the language specific keypad panel 14 keys 49 M1178202 FRU Keypad for Non Touchscreen model ZH B650 Includes the membrane keypad and the language specific keypad panel 14 keys FRU Itempart number FRU Item description FRU...

Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...

Страница 225: ...nd the connected devices are intact properly connected and secured to the right connectors The modules are properly connected and locked The pivoting module frame and battery door are properly locked Electrical safety tests Manufacturing week and year of the device Is there less than 12 months since the device was manufactured Yes Yes you do not have to perform the electrical safety tests To conti...

Страница 226: ...ed Polarity Normal 500 µA EN IEC 300 µA UL Ground open Polarity Normal 500 µA EN IEC 300 µA UL Ground open Polarity Reversed 500 µA EN IEC 300 µA UL Ground closed Polarity Reversed 500 µA EN IEC 300 µA UL Normal Condition NC Polarity Reversed 100 µA Module type S N Observed result Acceptance criteria PASS N A FAIL ECG SpO2 InvP Patient source leakage current tests using a test body Normal Conditio...

Страница 227: ...y Display Picture quality Text is readable images are clear and brightness is good Display Touchscreen control Touchscreen is correctly calibrated and operates correctly Device Information Device information is correct Configuration Information Monitor is correctly configured Keypad and remote Hard key Selected menu opens and selected activity starts Keypad and remote Trim Knob Selected menu opens...

Страница 228: ...ters Test page is printed to the selected printer Insite with EXC The patient monitor is active on Insite with EXC back office Test completion Notes Used service parts Signature Date Functional check Observed result Acceptance criteria PASS N A FAIL ...

Страница 229: ...damaged The monitor and the connected peripheral devices are properly mounted with specified mounting solutions The cables between the patient monitor and the connected devices are intact properly connected and secured to the right connectors The modules are properly connected and locked The pivoting module frame and battery door are properly locked Electrical safety tests Power outlet is correctl...

Страница 230: ...tarts up normally Display Picture quality Text is readable images are clear and brightness is good Display Touchscreen control Touchscreen is correctly calibrated and operates correctly PSM PDM identification Hemodynamic parameter data appears on the screen correctly E module identification E module parameter data appears on the screen correctly Keypad and remote Hard key Selected menu opens and s...

Страница 231: ...th EXC back office Recorder The header line contains the correct information The grid is clear The waveforms labels appear in the printout as configured Synchronization connector test ECG Arterial BP and Marker Out signals are correct Test completion Notes Used service parts Signature Date Functional check Observed result Acceptance criteria PASS N A FAIL ...

Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...

Страница 233: ... Clinical Information Center with full disclosure license NOTE CIC with full disclosure license is optional It is needed for documentation and reporting purposes only It enables you to view afterwards potential network connectivity issues between the transport monitor and the wireless MC Network and to print reports about waveform loss situations Wired CARESCAPE monitor the stationary monitor A CA...

Страница 234: ...unt the following aspects when preparing a test plan Identify areas with known or obvious low signal strength Identify areas with known sources of radio frequency interference causing high noise floor and or poor signal to noise ratio Identify the special characteristics in the building layout floors wings patient rooms and construction material used Identify the time and areas of congestion with ...

Страница 235: ...ythm Heart rate 80 bpm Amplitude 1 mV Configuring the patient monitor 5 Select ECG1 ECG2 and ECG3 waveform fields to the screen with adequate priority 6 In the Parameter Setup ECG Setup select ECG1 lead II ECG2 lead V1 ECG3 lead aVL 7 Start a patient case admit a patient in Data Pages Admit Discharge menu Wireless CARESCAPE monitor the transport monitor NOTE The PDM PSM module ECG cables and the p...

Страница 236: ...sh the web browser and to see up to date WLAN diagnostics data Test execution Execute the test procedure according to the test plan Contact the nursing staff to ensure access to the needed areas before starting the test 1 Move the roll cart to the starting point of the planned test route 2 Stop at each test point and perform the following tasks a On the transport monitor Verify that ECG waveforms ...

Страница 237: ... waveform loss is normal during roaming If longer or repeating waveform losses occur between test points make this an extra test point and report it the same way as observations in step 2 Summary and Reporting Include the following documents to the test results 1 Print the full disclosure reports from the CIC about the observed waveform loss situations 2 Print the snapshots of the WLAN diagnostics...

Страница 238: ...CARESCAPE Monitor B650 C 6 6 2062973 002 Test Form Observations Test summary and recommended actions Test point Time hh mm ss RSSI dBm Transmit rate Mbps Observations Date time of testing Signature ...

Страница 239: ......

Страница 240: ...ms Information Technologies Inc 8200 West Tower Avenue Milwaukee WI 53223 USA Tel 1 414 355 5000 1 800 558 5120 US only Fax 1 414 355 3790 Asia Headquarters GE Medical Systems Information Technologies Asia GE China Co Ltd No1 Huatuo Road Zhangjiang Hi tech Park Pudong Shanghai P R China 201203 Tel 86 21 5257 4650 Fax 86 21 5208 2008 www gehealthcare com ...