H3-38
006-252-02
10.22.99 FM2569
FM-DP L
Service Manual
H3-38
MAINTENANCE 3. TROUBLESHOOTING
Error code
Error name
Significance
Probable cause and remedy
Detailed
information format
0613
DSP self-diagnosis time-
out error
The self-diagnosis of the DSP did not
end within the specified time (90
seconds).
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
0614
DSP image processing
2 parameter setting error
Failed in writing the image processing 2
parameter in the DSP external memory.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
0615
DSP image processing
2 start time-out error
The DSP image processing 2 start
process did not end within the specified
time (90 seconds).
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
0616
DSP image processing
2 start error
The DSP image processing 2 start
results were faulty.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the
MTH board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
0617
DSP image processing
2 start error
The DSP image processing 2 start
results were faulty. (DSP4)
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the
MTH board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
0618
DSP image processing
2 start error
The DSP image processing 2 start
results were faulty. (DSP5)
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, and image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
6-C
Содержание FM-DP L
Страница 1: ...FM DP L Service Manual MAINTENANCE...
Страница 31: ...H1 29 006 252 02 10 22 99 FM2569 FM DP L Service Manual H1 29 MAINTENANCE 1 OUTLINE OF EQUIPMENT BLANK PAGE...
Страница 40: ...H3 7 2 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 2 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 41: ...H3 7 3 MAINTENANCE 3 TROUBLESHOOTING 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 3 BLANK PAGE...
Страница 42: ...H3 7 4 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 4 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 56: ...H3 19 2 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 2 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 57: ...H3 19 3 MAINTENANCE 3 TROUBLESHOOTING 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 3 BLANK PAGE...
Страница 58: ...H3 19 4 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 4 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 152: ...H3 89 16 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 89 16 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 244: ...H3 181 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 181 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 248: ...H3 185 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 185 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 252: ...H3 189 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 189 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 340: ...H3 277 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 277 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 349: ...FM DP L Service Manual PREVENTIVE MAINTENANCE...