H3-88
MAINTENANCE 3. TROUBLESHOOTING
Error code
Error name
Significance
Probable cause and remedy
Detailed
information format
2n3B
Film for changing
already output
(Image information
change)
Logic conflict occurred once with host
Continuable
None
2n3C
No print ID for changing
print information
Logic conflict occurred once with host
Continuable
None
2n3D
Film for changing
already output
(Print information
change)
Logic conflict occurred once with host
Continuable
None
2n3E
Cannot set as board
not for setting
Incorrect system setting
Continuable/set normal settings after continuation
None
2n50
No final storage image
after load mode ON
Incorrect operation
Continuable
None
2n51
Image copy error during
storage when load
mode ON
Insufficient memory
Continuable
None
2n52
No load mode ON
image during storage
when load mode ON
Incorrect operation
Continuable
None
2n80
HD image write error
Faulty HDD
Faulty MTH board
Replace the MTH board/HDD
None
2n81
HD overlay write error
Faulty HDD
Faulty MTH board
Replace the MTH board/HDD
None
2n85
HD directory information
write error
Faulty HDD
Faulty MTH board
Replace the MTH board/HDD
None
2n86
HD image information
write error
Faulty HDD
Faulty MTH board
Replace the MTH board/HDD
None
2n87
HD print information
write error
Faulty HDD
Faulty MTH board
Replace the MTH board/HDD
None
20F0
Error during SIO
transmission
Error occurred in the SIO during
exchange with the density
measurement unit.
Defect of SIO of the MTH board.
1. Replace the MTH board.
None
20F1
BCC error during SIO
reception
BCC error occurred during exchange
with the density measurement unit.
Defect of communication unit of the density measurement unit,
defect of communication cable, defect of SIO of the MTH
board.
1. Replace the density measurement unit.
2. Replace the communication cable.
3. Replace the MTH board.
None
20F2
Insufficient data error
during SIO reception
Data ran out during exchange with the
density measurement unit.
Defect of communication cable, defect of SIO of the MTH
board, defect of communication unit of the density
measurement unit.
1. Replace the communication cable.
2. Replace the MTH board.
3. Replace the density measurement unit.
None
Содержание FM-DP L
Страница 1: ...FM DP L Service Manual MAINTENANCE...
Страница 31: ...H1 29 006 252 02 10 22 99 FM2569 FM DP L Service Manual H1 29 MAINTENANCE 1 OUTLINE OF EQUIPMENT BLANK PAGE...
Страница 40: ...H3 7 2 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 2 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 41: ...H3 7 3 MAINTENANCE 3 TROUBLESHOOTING 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 3 BLANK PAGE...
Страница 42: ...H3 7 4 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 7 4 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 56: ...H3 19 2 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 2 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 57: ...H3 19 3 MAINTENANCE 3 TROUBLESHOOTING 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 3 BLANK PAGE...
Страница 58: ...H3 19 4 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 19 4 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 152: ...H3 89 16 006 252 05 06 30 01 FM3040 FM DP L Service Manual H3 89 16 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 244: ...H3 181 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 181 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 248: ...H3 185 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 185 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 252: ...H3 189 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 189 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 340: ...H3 277 006 252 02 10 22 99 FM2569 FM DP L Service Manual H3 277 MAINTENANCE 3 TROUBLESHOOTING BLANK PAGE...
Страница 349: ...FM DP L Service Manual PREVENTIVE MAINTENANCE...