[ 1-- INTERFACE ]
Interface Block
The interface main block is shown below.
[
1-- INTERFACE
]
INTERFACE MAIN BLOCK
Press
key sequentially to access this main block,
or press key or
key to enter configuration parameters,
or press
key to go to the next block,
or press
key to exit from programming.
[
11- RS232C
]
RS 232C SERIAL PORT
Press key or
key again to enter this menu.
Or press
key to go to the next sub-block.
[
111 FORMAT
:
PRN
T
]
Data format of the serial port
NO
CONT
FAST
PRNT
BSI
MBHL
MBLH
:
:
:
:
:
:
:
Port is disabled
Continuous data output
High speed continuous data output
Printout
BSI format for PC, PLC interface
Modbus RTU High-Low format
Modbus RTU Low-High format
Press
key to select output format different than Flintec
continuous data format while the selection is CONT in the
information display. Available functions are seen on the weight
display after pressing
key sequentially. Press
key after
selecting the data format.
Refer to Appendix 5 for details data structure.
[
FLINTEC
]
FLINTEC
HBM
TOLEDO
SYSTEC
SMA
SARTOR
RINSTR
AVERY
BASTER
LM2
:
:
:
:
:
:
:
:
:
Flintec continuous format.
Commonly used by HBM, GSE, PT, Systec, Rinstrum.
Commonly used by Toledo, Mettler Toledo.
Commonly used by Systec.
Commonly used by USA producers, Cardinal, Rice Lake etc.
Commonly used by Sartorious.
Commonly used by PT, Rinstrum, HBM, GSE.
Commonly used by Avery E1205.
Commonly used by Baster.
Flintec LM2 (BX1 par.000=6)
[
112 BAUD
: 9600 ]
Baud rate
1200
2400
4800
9600
:
:
:
:
1200 19K2 : 19200
2400 38K4 : 38400
4800 57K6 : 57600
9600
[
113 LENGHT
:
8 ]
Data Length
7
8
:
:
7 bits
8 bits
[
114 PARITY
:
NO ]
Parity
NO
ODD
EVEN
:
:
:
No parity
Odd parity
Even parity
[
115 CSUM
:
NO ]
Checksum at continuous and BSI formats.
NO
YES
:
:
Checksum disable
Checksum enable