FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
98
of
146
Modbus Command Table;
Address R/W Word
Command
Definition
40001
R
2
Actual weight (Net if the indication is in Net, Peak value or Hold value)
40003
R/W
2
Tare weight
40005
R
2
Gross weight
40007
R
2
Status
Bit
Definition
B0
0
– Dynamic is inactive
1
– Dyn weight is calculating
B1
0
– Weight is actual w.
1
– Weight is dynamic result
B2
0
– Weight is Stable
1
– Weight is unstable
B3
0
– Gross mode
1
– Net mode
B4
Not used
B5
0
– Not power on zeroed
1
– Zeroed with pwr on zero
B6
0
– First Unit (power on)
1
– Second Unit
B7-B10
Not used
B11
0
– Key lock is passive
1
– Key lock is active
B12
0
– Out of center of zero
1
– Weight is in center of zero
B13
0
– High res. is passive
1
– High res. is active
B14
0
– Basic Peak is passive
1
– Basic Peak is active
B15
0
– Hold is passive
1
– Hold is active
B16-B26
Not used
B27
0
– None
1
– Decimal point is X.XXXX
B28
0
– None
1
– Decimal point is X.XXX
B29
0
– None
1
– Decimal point is X.XX
B30
0
– None
1
– Decimal point is X.X
B31
0
– None
1
– No decimal point
40009
R
2
Error
Status
Bit
Definition
B0
0
– No Error
1
– Low voltage det.
B1
0
– No Error
1
– In programming mode
B2
0
– No Error
1
– System error
B3
0
– No Error
1
– ADC under range
B4
0
– No Error
1
– ADC over range
B5
0
– No Error
1
– ADC out of range
B6
0
– No Error
1
– Tilt Switch is active
B7
– B31
Not used
40011
R
2
Heartbeat for connection checking, this value is increased every
100 milliseconds.
40013
R
2
Last print value
40015
R/W
2
CN (Label number)
Refer to parameter [254] on page 56
40017-24
R
8
40025
R/W
2
Commands
Dec
Definition
0
None
1
Zero
2
Tare
3
Clear
4
5
Reprint the last label
6
High resolution enable
7
High resolution disable
8
Unit change (from first to second unit)
9
Unit change (from second to first unit)