FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
9
of
146
RS
4
22
Baud rate
1200 to 57600 programmable
Data
Length 7 or 8 bits; parity even, odd or none
Stations
Up to10 stations
E
the
rne
t TC
P
/I
P
Transmission rate
10 / 100 Mbit/s, Full duplex
TCP/IP settings
Manual IP assign over Indface2x PC Software or by keys in programming mode
Connection method Server or Client
Installation
Switched Ethernet transmission with shielded twisted pair cables and RJ-45.
Isolation
Galvanically isolated bus electronics
Response speed
Up to 4 ms response delay after read/write commands
US
B
Connection
Standard USB Micro-B cable
Response speed
Min. 4 ms response delay after read/write commands
Digital Inputs and Outputs (opional)
Digital Inputs
Optoisolated 4 digital input, 12 to 28 VDC, 10mA.
Any input(s) can be used as a Remote input of your PLC over BSI or Modbus.
Digital Outputs
5 free relay contact, 250 VAC or 30 VDC, 0.2A
Any output(s) can be used as a Remote output of your PLC over BSI or Modbus.
Analogue Output
(
opional)
Voltage output
0-5 VDC, 0-10 VDC
Current output
4-20mA, 0-20mA
Resolution
60 000 steps
Max. cable length
300 meters
Max. load resistance
(current output)
500 Ω
Minimum load
resistance (voltage
output)
10 kΩ
Memory
Application memories
Set Memory to save limit values at your industrial weighing.
Memory size is 500 item records
Alibi memory (optional) 99 999 records
Communication
Connectable with
PC, PLC, Printer, Remote display, EPL printer etc.
RS
2
32
Isolation
Galvanically isolated.
Baud rate
1200 to 57600 programmable
Data
Length 7 or 8 bits; parity even, odd or none
RS
4
85
Isolation
Galvanically isolated.
Baud rate
1200 to 57600 programmable
Data
Length 7 or 8 bits; parity even, odd or none
Stations
Up to 31 stations