FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
41
of
146
6 P
ROGRAMMING AND
C
ALIBRATION
You will find the programming and calibration procedure of FT-111 weighing indicator in this section. The arrow on
navigation keys indicates the function of the keys in programming menu. The basic meanings of these keys are;
A set-up DIP switch is located near the load cell connector at the FT-111 rear side as shown in the figure below and its
position should be “ON” ( downward ) to change the metrological related parameters including calibration.
Figure 6.1- The location of calibration DIP switch.
6.1 Entering to the Programming and Calibration
Enter the programming menu as described below. Legally related parameters can be changed if the calibration
dip switch is at ON position. These parameters are marked with
M
symbol in the parameter table in this section.
If you will change any legally relate parameter or will perform calibration, set the calibration switch to the ON
position before entering to the programming.
Display
Operation
[123.456 kg]
Press
key until [
PASSWORD :
] prompts seen.
[PASSWORD]
+
+
+
keys sequentially for service entry.
[**** ]
Press
key.
[1
—INTERFACE ]
You entered to the programming main menu and the first main
block [1
—INTERFACE ] prompt appears.
Navigation
between
parameters
Exit from any
entry or block
or from
programming.
Go to
previous block
or parameter.
Next block.
Previous
block.
Go to
next block
or parameter.
Enter the
parameter to
edit it.
Parameters’
value
Exit with
previous value
Move to the
left digit.
Next option or
Increase the
blanking digit
Previous
option or
Decrease the
blanking digit
Move to the
right digit.
Confirm and go
to the
next parameter.
DIP Switch
Description
1
Calibration SW
OFF : Locked
ON : Set- up
2
Reserve