FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
51
of
146
[
177 LF AFTE
:
+4 ]
Line feed after printout.
:
:
+ = Forward , - = Backward
NO,1,2….9 : Line feed quantity after data.
Example: -2 means 2 line feed backward.
[
178 FORM FE
:
NO ]
Form feed.
NO
YES
:
:
Disable
Enable
[
179 LEFT SP
:
3 ]
Space from left of the label.
0…9
[
17A COPY
:
1 ]
Copy quantity.
1…9
[
19- ANALOG OUT
]
ANALOGUE OUTPUT
(Only FT-111 AN)
Press key or
key again to enter this menu.
Press
key to go to beginning of the sub-block or press
key to go to the next main-block.
[
191 TYPE
:
4-20 ]
Analog output type
4-20
0-20
0-10
0-5
:
:
:
:
4 mA - 20 mA
0 mA - 20 mA
0 VDC- 10 VDC
0 VDC- 5 VDC
[
192 MINIMUM
:
00.0 ]
The minimum of the analogue output. Default 00.0 means the
minimum output is not drifted.
e.g. enter 1.0 to set output to 1.0 volt at 0 kg in 0 -10 V range.
[
193 MAXIMUM
:
00.0 ]
The maximum of the analogue output. 00.0 means the maximum
output is not drifted.
e.g. enter 9.0 to set output to 9.0 volt at Max load in 0 -10 V range.
[
194 SOURCE
:
GRO
S
]
Source of the analogue output
GRO
S
INDI
:
:
Gross weight
Indicated weight
[
195 ZERO ADJ :
]
Displays the count value of unloaded scale.
Increase or decrease by pressing
or
; or press key to
enter the value by related keys to change the unloaded scale
output.
[
196 SPAN ADJ :
]
Displays the count value of the full loaded scale.
Increase or decrease by pressing
or
; or press key to
enter the value by related keys to change the analogue output gain.
[
197 AUTO ADJ :
NO ]
Set analogue output to calibration range
NO
YES
:
:
No
Automatic adjustment between min and max
limits if changed.