FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
40
of
146
Search with Alibi record number:
1. Press
key after selecting NUM in parameter 813.
2. Press numerical keys to enter alibi number in the printout data and press
key. You
may navigate in the alibi memory with
or
keys after entering alibi number.
3. The weighing data in the alibi memory is indicated on the display.
4. Press
key to print this record and the previous nine weighing.
5. Press
key 4 times to return operation.
Search with date:
1. Press
key after selecting DATE in parameter 813.
2. Press numerical keys to enter date in the printout data and press
key. The date format
entry should be as in the operation. e.g. date entry should be DDMMYY if parameter 251 is
set to DMY. You may navigate in the alibi memory with
or
keys after entering
the date.
3. The weighing data in the alibi memory is indicated on the display.
4. Press
key to print this record and the previous nine weighing.
5. Press
key 4 times to return operation.
Search with weight value:
1. Press
key after selecting NET, gross or tare which is suitable in parameter 813.
2. Press numerical keys to enter weight value in the printout data and press
key.
3. The alibi record is indicated on the display.
4. Press
key to print this record and the previous nine weighing.
5. Press
key 4 times to return operation.
Transfer all Alibi records
1. Go to the parameter [ 8-- METROLOGY ] main block in set up.
2. Press key 5 times to access parameter 814.
3. Press
or
keys to select PRNT
and press
key to start transferring.
4. Or you may stop transferring by pressing
key.
About Alibi Memory
1. Go to the parameter [ 8-- METROLOGY ] main block in set up.
2. Press key 6 times to access parameter 815.
3. Press
key to print alibi information.
Format Alibi Memory
Warning : Alibi memory formatting should be done by authorized person. Otherwise erasing alibi memory may
bring you undesired legal responsibility.
You may need to format alibi memory after installing the used SD card, to erase previous records before starting
usage in legal or to erase alibi memory if you don’t need previous records legally. This process runs if the
calibration switch is at programming position.
1. Go to the parameter [ 8-- METROLOGY ] main block in set up.
2. Press key 7 times to access parameter 816.
3. Select YES and press
key.
4. The warning prompt [ ARE YOU SURE? ] appears on the display.
5. Press
key to start formatting or Press
key to escape.
6. Press
key 3 times to return operation.
The empty fields are indicated as [ ------- ] and transferred as [ NO RECORD ]; and the corrupted records
are indicated as [ Error ] and transferred as [ CORRUPTED ].