8.2 Digital outputs
Digital outputs can be programmable as free setpoints which can be programmed for different functions as seen
below. The output (s) can be produced as threshold or window additional to the well-known basic output at free
programmable setpoints.
S AIN Free setpoint of absolute indicated weight
The status of the digital output changes with comparing the set point value and absolute indicated weight
value as seen below.
As an example,
if SP1
SP1
LOW
SP1
HIGH
= 100 kg or
= 100 kg and
= 150 kg.
BAS-HI
Basic, active high
Output is high if absolute indicated weight is
heavier than setpoint.
( Default )
BAS-LO
Basic, active low
Output is low if absolute indicated weight is
heavier than setpoint.
THR-HI
Threshold, active high
Output goes to high if absolute indicated
weight is heavier than setpoint high. Returns
the low if the weight is less than setpoint low.
THR-LOW
Threshold, active low
Output goes to low if absolute indicated weight
is heavier than setpoint high. Returns the high
if the weight is less than setpoint low.
VIN-HI
Window, active high
Output goes to high if absolute indicated
weight is between setpoint high and setpoint
low.