Standard calculation method
Regardless of the shape of the tank, average temperature is calculated using the following
formula:
Formula: (T1 + T2 + T3) / Number of elements in liquid phase = Average temperature
(3.5 °C (38.3 °F) + 3.0 °C (37.4 °F) + 2.0 °C (35.6 °F)) / 3 = 2.83 °C (37.1 °F)
1
2
3
4
5
A0038546
32
Standard calculation method for liquid temperature
1
Element No.5: 4.5 °C (40.1 °F) (T5)
2
Element No.4: 4.0 °C (39.2 °F) (T4)
3
Element No.3: 2.0 °C (35.6 °F) (T3)
4
Element No.2: 3.0 °C (37.4 °F) (T2)
5
Element No.1: 3.5 °C (38.3 °F) (T1)
Advanced calculation method
Average temperature is calculated by adding a corrective factor for unequal volume
distribution.
Formula: (T1*V1 + T2*V2 + T3*V3) / (V1 + V2 + V3) = Average temperature
Parameters that are related to V = additional volume factors are determined in VH53,
54 and 55.