Operation
Prothermo NMT532
54
Hauser
8.3.8
Temperature adjustment: VH70-VH79
Code
Display
Details
VH70
Element Select
(Element number
assignment)
Item type
Selection
Range
0 to 19
The element number for performing temperature adjustments is
selected (Element 0-15 = Element 1-16, 19 = Reference 100 Ω
resistance).
VH71
Zero Adjust
(Zero adjustment of
temperature
measurement
element)
Item type
Read / Write
Default
0
Range
–1
000.0 to 1
000.0
Performs zero adjustment for individual elements that were selected in
VH70. Unlike a standard temperature device, the reading value can be
adjusted when the measured temperature shows minor correction
values.
If element No.2 displays 25.4 °C (77.72 °F) and a standard
temperature device displays 25.2 °C (77.36 °F), the matrix is set
to -0.2.Once set, the correction value of element No.2 based on
the actual measured value will be –0.2 °C (31.6 °F).
VH72
Adjust Span
(Adjustment of
temperature
measurement
element span)
Item type
Read / Write
Default
1
Range
0.8 to 1.2
Span adjustment is applied to all installed temperature measurement
elements. This correction value is multiplied by the actual measured
values.
VH73
Temperature X
(Temperature X)
Item type
Read only
Temperature of the elements that were selected in VH70. Also displays
each temperature measurement element that was displayed in VH10-
VH25. The value is calculated based on the following formula:
VH73: Temperature X = Unadjusted temperature x span (VH72) + Zero
offset (VH71)
VH74
Position X
(Element position)
Item type
Read / Write
Range
0 to 99
999 mm
Position of the elements that were selected in VH70. If "Unequal" was
selected in VH85, the position of each element can be set here.
VH75
Resistance X
(Element resistance)
Item type
Read only
Shows the measured resistance for elements that were selected in
VH70.
VH76
Resistance Adj.
(Element resistance
adjustment)
Item type
Read / Write
Default
0
Range
–1
000.0 to 1
000.0
Adjusts the resistance of elements selected in VH70. Minor resistance
adjustments can be applied on the reading value.
Under identical environmental conditions, if the selected element
No.5 displays 100.3 Ω and a standard high-precision resistor
displays 100 Ω, -0.3 is set in this matrix. Once set, the correction
value of element No.5 based on the actual measured value will be
–0.3 Ω. Be careful when configuring this setting in VH76, as it
will apply to all elements.
VH77
Element Type
(Element type)
Item type
Selection
Selection
Pt100, Cu90, Cu100, PtCu100, JPt100