Prothermo NMT532
Operation
Hauser
59
8.4
WB (water bottom) level input from host
This function is available in V1.53 and later versions. If the specification does not have a
WB probe, VH50 WB can be entered manually from the HART Master.
This function is available for specifications that do not have a WB probe. It will not be
applied to multi-element calculations. If the WB probe error is resolved, the previous value
will be used. It is 0 mm when the power is turned off or on.
Process for entering 876.5 mm (34.51 in) (example) as a WB value:
1. Enter 129 in Device Specific Commands.
2. Enter 047 (0x047E/VH50) in Data (Hex).
047E indicates variable address 1150.
3. Enter float value 445B2000.
It will show the WB value 876.5 mm (34.51 in).
This completes the input process.
31 indicates millimeter (mm).
1
A0038557
38
Screen 1
1
Variable address
Process for entering 2
345.6 mm (92.35 in) (example) as a WB value:
1. Enter 145 in Device Specific Commands.
2. Enter 50 (VH50) in Data (Hex).
VH50 will be the matrix for NMT532, NMT532 and NCT530.
3. Enter the float value 4512999A.
It will show WB value 2
345.6 mm (92.35 in).
This completes the input process.
31 indicates millimeter (mm).