Prothermo NMT532
Operation
Hauser
55
Code
Display
Details
The element conversion formula is selected for when another brand's
average temperature probe is connected to the NMT532.
L
CAUTION
Changing the parameters:
NMT532 is comprised of spot, element array and PT100 element
types.
‣
Changing the parameters may cause erroneous calculations or
unnecessary error displays.
Element conversion
formula
Pt100 (formula above 0 °C):
R = -0.580195 x 10-4 x T2 + 0.390802 x T +100
Pt100 (formula below 0 °C): R = -4.2735 x 10-10 x T4 + 4.273 x 10- 8
x T³ - 0.58019 x 10- 4 x T² + 3.90802 x T + 100
Cu90: R = 0.3809 x T + 90.4778
Cu100: R = 0.38826 x T + 90.2935
PtCu100: R = 3.3367 x 10-7 x T3 - 2.25225 x 10-5 x T2 +0.38416 x T
+ 100.17
R: Resistance, T: Temperature
VH78
Average Number
(Sampling number)
Item type
Read / Write
Default
1
Range
1 to 10
The number of resistance samplings from all temperature
measurement elements can be changed, including that of the reference
resistance installed on the circuit of the main unit.
Increasing the sampling number will allow for a more accurate
measurement, but it will slow down the device's overall scan
time. Element selection frequency: Approx. 2 seconds/element,
sampling maximum element number 21 (Number of elements:
16, Internal reference resistance: 5)
VH79
Protect Code
(Access code)
Item type
Read / Write
Default
0
Range
0 to 999
Access code 530 enables selection and writing.
8.3.9
Device setting 1: VH80-VH89
Code
Display
Details
VH80
Present Error
(Error
information)
Item type
Read only
Screen where error information is displayed. The following error codes will be
displayed. For details, refer to "Troubleshooting" .
Error
code
0
No error presence
1
Common line open
3
#1 element open
4
#1 element short
5
#2 element open
6
#2 element short
7
#3 element open
8
#3 element short
9
#4 element open
10
#4 element short