Adjustment and settings
Prothermo NMT532
42
Hauser
Access code
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Access code (14714)
Conditions
Configure device? = Yes
Description
Displays the access code.
Input range
0-65535
Factory setting
0
Additional information
Read access: Operator
Write access: Maintenance
Total number of elements
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Total No. elements (14730)
Description
Displays the total number of elements that can be configured.
Additional information
Read access: Operator
Write access: -
Bottom point
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Bottom point (14729)
Description
Displays the bottom-point temperature element.
Input unit
Numerical value (mm)
Factory setting
0 mm
Additional information
Read access: Operator
Write access: Maintenance
Temp element short
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART device configuration → Temp elem. short (14731)
Description
Configures the error code for when an element short-circuits.
Input unit
Numerical value (
℃
)
Factory setting
0 °C
Additional information
Read access: Operator
Write access: Maintenance