Operation
Prothermo NMT532
62
Hauser
8.6
Write-protection switch (write-protection plug)
When the write-protection switch in software version 1.53 or later is used, all parameters
become write protected. In PTB specifications, this comes with the product.
Do not disconnect and reconnect while the power is on.
W & M
20 (0.79) ± 3 (0.12)
a
A0038560
41
Write-protection switch
a
To NMT532 CPU CN3
8.7
Module configuration
After replacing the electric housing module, parameters must be re-entered manually into
the electric housing module to maintain proper operation. Check the following matrix
parameters after replacing the electric housing module.
NMS5/NMS7 GVH
FieldCare / Details
443
Level select
450-459
Element position No. 1-9
470
Select point (elements 0-15)
474
Position X (element position that was set in GVH=470)
482
Element number
485
Kind of interval
486
Bottom point
487
Element interval (when selecting equal interval at GVH=485)