Prothermo NMT532
Operation
Hauser
61
2
3
4
1
1000 (39.37)
500 (19.69)
1500 (59.06)
a
b
A0038559
40
Example of temperature elements near the tank bottom
a
Temperature element 1
b
Temperature element 2
1
Case 1
2
Case 2
3
Case 3
4
Case 4
• Case 1: If the liquid level is within this range, temperature element a will not be
used in average liquid temperature calculations.
• Cases 2 and 3: If the liquid level is within this range, temperature element a will be
used in average liquid temperature calculations.
• Case 4: If the liquid level is within this range, none of the temperature elements will
be used in average liquid temperature calculations.
8.5.1
Temperature adjustment: VH92
The following table shows the settings.
VH92
setting
Presence
of liquid in
tank
Temp.
element
open
(disrupted)
Temp.
element
short
(short
circuit)
Average liquid temperature error output
ON
No
No
No
358 °C (676.4 °F)
No
Yes
No
358 °C (676.4 °F)
No
No
Yes
358 °C (676.4 °F)
Yes
No
No
Average liquid temperature
Yes
Yes
No
Open error set value (default 359 °C (678.2 °F))
Yes
No
Yes
Short circuit error set value (default –49.5 °C (–57.1 °F))
OFF
(Default)
No
No/Yes
No/Yes
Average gas temperature
(The malfunctioning temperature element is skipped)
Yes
No/Yes
No/Yes
Average liquid temperature
(The malfunctioning temperature element is skipped)