Prothermo NMT532
Adjustment and settings
Hauser
43
Temp element open
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART device configuration → Temp elem. open (14732)
Description
Configures the error code for when an element is open.
Input unit
Numerical value (
℃
)
Factory setting
0 °C
Additional information
Read access: Operator
Write access: Maintenance
Output error
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART device configuration → Output error (14733)
Description
Selects the error display for when an element shorts and opens.
Selection
OFF
ON
Additional information
Read access: Operator
Write access: Maintenance
Gain adjust
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART device configuration → Gain adjust (14736)
Description
Adjusts the temperature of all elements and references 0 and 17.
Input unit
Numerical value
Factory setting
0
Additional information
Read access: Operator
Write access: Maintenance
Kind of interval
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Adjust interval (14744)
Description
Sets the type of element interval.
Selection
Equal
Unequal
Factory setting
Equal
Additional information
• Equal division: Bottom point + Element interval
• Unequal: Set manually
Read access: Operator
Write access: Maintenance
Содержание Prothermo NMT532