Prothermo NMT532
Operation
Hauser
47
Code
Display
Details
(Liquid level)
Range
0 to 99
999 mm
Displays the liquid level inside the tank as configured by a level gauge.
If a level gauge is not connected, a directly input liquid level can be
used as a device test.
VH07
Temperature 0
(Element 0
temperature)
Item type
Read only
Tolerance
–1.0 to 1.1 °C (30.2 to 33.8 °F)
Checks that the measuring temperature resistor temperature
conversion is executed correctly.
VH09
Temperature 17
(Element 17
temperature)
Item type
Read only
This temperature is used for checking when shipping from the factory.
8.3.2
Temperature measurement elements 1: VH10 to VH15 (VH16
to 19 are not available in NMT 532)
Code
Display
Details
VH10-19
Temperature 1-10
(Temperature of
elements 1 to 10)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Displays individual temperature measurement elements.
8.3.3
Temperature measurement elements 2: VH20 to VH25 (VH26
to 29 are not available in NMT532)
Code
Display
Details
VH20-25
Temperature 11-16
(Temperature of
elements 11 to 16)
Item type
Read only
Range
–200 to 240 °C (–328 to 464 °F)
Displays individual temperature measurement elements.
VH26
Selec. Ave Method
(Average
temperature
calculation method)
Item type
Selection
Selection
Standard / Advanced
Selects the method of average temperature calculation.