Technical data
iTEMP TMT85
50
Hauser
13
Technical data
13.1 Input
Measured variable
Temperature (temperature-linear transmission behavior), resistance and voltage.
Measuring range
Two independent sensors can be connected. The measuring inputs are not galvanically
isolated from each other.
Resistance thermometer
(RTD) as per standard
Designation
α
Measuring range limits
IEC 60751:2008
Pt100 (1)
Pt200 (2)
Pt500 (3)
Pt1000 (4)
0.003851
–200 to +850 °C (–328 to +1
562 °F)
–200 to +850 °C (–328 to +1
562 °F)
–200 to +250 °C (–328 to +482 °F)
–200 to +250 °C (–328 to +482 °F)
JIS C1604:1984
Pt100 (5)
0.003916
–200 to +649 °C (–328 to +1
200 °F)
DIN 43760 IPTS-68
Ni100 (6)
Ni1000
0.006180
–60 to +250 °C (–76 to +482 °F)
–60 to +150 °C (–76 to +302 °F)
Edison Copper Winding No.
15
Cu10
0.004274
–100 to +260 °C (–148 to +500 °F)
Edison Curve
Ni120
0.006720
–70 to +270 °C (–94 to +518 °F)
GOST 6651-94
Pt50 (8)
Pt100 (9)
0.003910
–200 to +1
100 °C (–328 to +2
012 °F)
–200 to +850 °C (–328 to +1
562 °F)
OIML R84: 2003
GOST 6651-2009
Cu50 (10)
Cu100 (11)
0.004280
–200 to +200 °C (–328 to +392 °F)
-
Pt100 (Callendar van Dusen)
Nickel polynomial
Copper polynomial
-
10 to 400 Ω, 10 to 2
000 Ω
10 to 400 Ω, 10 to 2
000 Ω
10 to 400 Ω, 10 to 2
000 Ω
• Type of connection: 2-wire, 3-wire or 4-wire connection, sensor current: ≤ 0.3 mA
• With 2-wire circuit, compensation of wire resistance possible (0 to 30 Ω )
• With 3-wire and 4-wire connection, sensor wire resistance up to max. 50 Ω per wire
Resistance transmitter
Resistance Ω
10 to 400 Ω
10 to 2
000 Ω
Thermocouples as
per standard
Designation
Measuring range limits
IEC 60584, Part 1
Type A (W5Re-W20Re) (30)
Type B (PtRh30-PtRh6) (31)
Type E (NiCr-CuNi) (34)
Type J (Fe-CuNi) (35)
Type K (NiCr-Ni) (36)
Type N (NiCrSi-NiSi) (37)
Type R (PtRh13-Pt) (38)
Type S (PtRh10-Pt) (39)
Type T (Cu-CuNi) (40)
0 to +2
500 °C (+32 to +4
532 °F)
+40 to +1
820 °C (+104 to +3
308 °F)
–270 to +1
000 °C (–454 to +1
832 °F)
–210 to +1
200 °C (–346 to +2
192 °F)
–270 to +1
372 °C (–454 to +2
501 °F)
–270 to +1
300 °C (–454 to +2
372 °F)
–50 to +1
768 °C (–58 to +3
214 °F)
–50 to +1
768 °C (–58 to +3
214 °F)
–260 to +400 °C (–436 to +752 °F)
Recommended temperature range:
0 to +2
500 °C (+32 to +4
532 °F)
+500 to +1
820 °C (+932 to +3
308 °F)
–150 to +1
000 °C (–238 to +1
832 °F)
–150 to +1
200 °C (–238 to +2
192 °F)
–150 to +1
200 °C (–238 to +2
192 °F)
–150 to +1
300 °C (–238 to +2
372 °F)
+50 to +1
768 °C (+122 to +3
214 °F)
+50 to +1
768 °C (+122 to +3
214 °F)
–150 to +400 °C (–238 to +752 °F)
IEC 60584, Part 1;
ASTM E988-96
Type C (W5Re-W26Re) (32)
0 to +2
315 °C (+32 to +4
199 °F)
0 to +2
000 °C (+32 to +3
632 °F)
ASTM E988-96
Type D (W3Re-W25Re) (33) 0 to +2
315 °C (+32 to +4
199 °F)
0 to +2
000 °C (+32 to +3
632 °F)
DIN 43710
Type L (Fe-CuNi) (41)
Type U (Cu-CuNi) (42)
–200 to +900 °C (–328 to +1
652 °F)
–200 to +600 °C (–328 to +1
112 °F)
–150 to +900 °C (–238 to +1
652 °F)
–150 to +600 °C (–238 to +1
112 °F)
GOST R8.585-2001 Type L (NiCr-CuNi) (43)
–200 to +800 °C (–328 to +1
472 °F)
–200 to +800 °C (+328 to +1
472 °F)