Commissioning
iTEMP TMT85
32
Hauser
• The files required for commissioning and network configuration can be obtained as
described on → 27.
• In the case of the FOUNDATION Fieldbus
TM
, the device is identified in the host or
configuration system by means of the device ID (DEVICE_ID). The DEVICE_ID is a
combination of the manufacturer ID, device type and device serial number. It is unique
and can never be assigned twice. The structure of the DEVICE_ID can be broken down as
follows:
DEVICE_ID = 452B4810CE-XXXXXXXXXXX
452B48 = Hauser
10CE = TMT85
XXXXXXXXXXX = Device serial number (11-digit)
• For a quick and reliable head transmitter configuration, a wide range of configuration
wizards are available to guide the user through the configuration of the most important
parameters of the Transducer Blocks. For this, refer to the operating manual of your
operating and configuration software.
The following wizards are available:
Configuration wizards
Name
Block
Description
Quick Setup
Sensor Transducer
Configuration of the sensor input with the data
relevant to the sensor.
Quick Setup
Display Transducer
Menu-guided configuration of the display unit.
Set to OOS mode
Resource, Sensor
Transducer, Display
Transducer, AdvDiagnostic
Transducer, AI, PID and
ISEL
Sets the individual block to the "Out Of Service"
mode
Set to auto mode
Resource, Sensor
Transducer, Display
Transducer, AdvDiagnostic
Transducer, AI, PID and
ISEL
Sets the individual block to the "Auto" mode
Restart
Resource
Restarts the device with different options as to
which particular parameters should be reset to the
factory settings.
Sensor drift monitoring
configuration
AdvDiagnostic Transducer Settings for drift or differential monitoring with 2
connected sensors.
Calc.wizard for 2-wire
compensation value
Sensor Transducer
Calculation of the conductor resistance for 2-wire
compensation.
Set all TRD to OOS mode
All Transducer Blocks
Sets all Transducer Blocks simultaneously to the
"Out Of Service" mode
Set all TRD to auto mode
All Transducer Blocks
Sets all Transducer Blocks simultaneously to the
"Auto" mode
Show recommended action
Resource
Displays the recommended action for the
diagnostic event currently pending.
Calibration wizards
User sensor trim
configuration
Sensor Transducer
Menu guidance for linear scaling ( slope)
for adapting the measuring point to the process
(→ 66).
Factory trim settings
Sensor Transducer
Resets scaling to the "factory standard trim"
(→ 66).
RTD-Platin configuration
(Call.-Van Dusen)
Sensor Transducer
Entry of Callendar-Van-Dusen coefficients.
RTD-Copper configuration
Sensor Transducer
Entry of coefficients for nickel polynomial.
RTD-Nickel configuration
Sensor Transducer
Entry of coefficients for copper polynomial.
Содержание iTEMP TMT85