iTEMP TMT85
Diagnostics and troubleshooting
Hauser
45
Symptoms
Cause
Action/remedy
Sensor, sensing element
defective
Check the sensor, sensing element
RTD connection incorrect
Connect the connecting cables correctly (see
the "Electrical connection" section )
Programming
Incorrect sensor type set in the
SENSOR_TYPE
device function. Set the correct sensor type
Device defective
Replace device
9.3.2
Application errors for TC connection
Symptoms
Cause
Action/remedy
Measured value is incorrect/
inaccurate
Incorrect sensor orientation
Install the sensor correctly
Heat conducted by sensor
Observe the face-to-face length of the
sensor
Device programming is
incorrect (scaling)
Change scaling
Incorrect thermocouple type
(TC) configured
Change the
SENSOR_TYPE
device
function
Incorrect reference measuring
point set
See Section 13
Offset incorrectly set
Check offset
Interference via the
thermocouple wire welded in
the thermowell (interference
voltage)
Use a sensor where the thermocouple wire
is not welded
Sensor connected incorrectly
Connect the connecting cables correctly
(see the "Electrical connection" section )
Sensor, sensing element
defective
Check the sensor, sensing element
Programming
Incorrect sensor type set in the
SENSOR_TYPE
device function; set the
correct thermocouple (TC)
Device defective
Replace device
9.4
Spare parts
Spare parts currently available for your product can be found online at:
http://www.products.endress.com/spareparts_consumables
, temperature transmitter:
TMT85. Always quote the serial number of the device when ordering spare parts!
Type
Order number
Adapter for DIN rail mounting, DIN rail clip as per IEC 60715
51000856
Standard - DIN securing set (2 screws and springs, 4 shaft lock-down rings, 1 plug for the
display interface)
71044061
US - M4 securing set (2 screws and 1 plug for the display interface)
71044062