Technical data
iTEMP TMT85
62
Hauser
Field housing
All field housings have an internal geometry in accordance with DIN EN 50446, form B
(flat face). Cable glands in the diagrams: M20x1.5
Maximum ambient temperatures for cable glands
Type
Temperature range
Polyamide cable gland ½" NPT, M20x1.5 (non-Ex)
–40 to +100 °C (–40 to 212 °F)
Polyamide cable gland M20x1.5 (for dust ignition-proof area)
–20 to +95 °C (–4 to 203 °F)
Brass cable gland ½" NPT, M20x1.5 (for dust ignition-proof area)
–20 to +130 °C (–4 to +266 °F)
Fieldbus connector (M12x1 PA, 7/8" FF)
–40 to +105 °C (–40 to +221 °F)
TA30A
Specification
107.5 (4.23)
68.5 (2.7)
28
(1.1)
78 (3.1)
15.5 (0.6)
A0009820
• Two cable entries
• Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
• Material: aluminum, polyester powder coated
Seals: silicone
• Cable entry glands: 1/2" NPT and M20x1.5
• Color of head: blue, RAL 5012
• Color of cap: gray, RAL 7035
• Weight: 330 g (11.64 oz)
TA30A with display window in cover
Specification
107.5 (4.23)
91.6 (3.61)
28
(1.1)
78 (3.1)
15.5 (0.6)
A0009821
• Two cable entries
• Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
• Material: aluminum, polyester powder coated
Seals: silicone
• Cable entry glands: 1/2" NPT and M20x1.5
• Color of head: blue, RAL 5012
• Color of cap: gray, RAL 7035
• Weight: 420 g (14.81 oz)