3.1 OVERVIEW
COMPASS for molbox
is an application program intended to automate the test and calibration process
to the extent supported by the available calibration hardware and appropriate for the DUT being tested.
COMPASS
supports automated operation of
DHI
molbloc/molbox
transfer standards but can also be
used with any type of flow measurement system from any manufacturer. This feature allows
COMPASS
to be
used systematically to provide consistent data formats and report generation for all of your flow calibrations.
COMPASS
sets up and maintains setup requirements on devices under test (DUTs) (see Section 8) and
testing procedures (see Section 9) to associate with DUTs. This information is recalled to run tests.
Tests are run using the hardware that has been configured (see Section 10) in
COMPASS
. While running
a test, data is acquired and stored in a delimited Data File (see Section 12). Within
COMPASS
, Data Files
may be viewed, used to generate plots and to generate reports which can be customized with user
editable templates (see Section 16).
COMPASS
operations follow conventional Windows protocol for file
management, menu and message formatting, graphics and editing.
To run a test, first a DUT Definition must be created (see Section 8) to define the DUT that will be tested
and a Test Definition must be created (see Section 9) to provide
COMPASS
with the exact testing
procedure for the DUT. The Hardware Definition, must be set up with the necessary instruments to
support the DUT, including the setup of the molbox remote interface. If the DUT requires voltage or
current measurement, a DMM must be set up using the Hardware Definition Tool (see Section 10).
The devices set up in the Hardware Definition must be selected as the current System Configuration
(see Section 10.2). Once a DUT and test have been set up, and the Hardware Definition has been
defined and selected, tests can be run (see Section 7.2).
The results of the test are stored in a Data File (see Section 18) that is designed to be used by
COMPASS
to generate reports or easily exported to other applications by the user for further analysis.
Over 40 individual data items are logged for each test point.
COMPASS
includes integrated analysis and
reporting functions. Fully customizable reports and plots can be created using Data Files (see Section 16).
Other than running tests,
COMPASS
supports direct molbox and DUT run options to allow remote control
of the molbox and/or a DUT with all
COMPASS
features active (see Sections 0, 7.4). This feature
provides a remote interface to all measurement and control instruments required by the DUT and setup in
the current System Configuration (see Section 10.2). The output information displays on various program
run screens. Use this feature to perform tests or operations that do not require creation of a Data File or
execution of a specific test script.
Many operational preferences relating to how tests are initialized and run, and how data is gathered and
stored are configurable with user options (see Section 11). The use of these configurable preferences
results in a customized test environment to meet a very wide variety of testing requirements. By selecting
the proper options, a user can start an automated test and return to a custom formatted report.
Detailed information and assistance on
COMPASS
application and functions are available in this manual
(see Sections 4, 5 and 20) and with on-line help (accessed by pressing
[F1]
or selecting
[Help]
).
3.1.1 BASICS
SUMMARY
In summary, the following steps and sections should be referred to when setting up and
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...