COMPASS
for molbox™ User’s Manual
© 1998-2003
DH Instruments, Inc
.
Page 38
6.6.8 FLOW
CONTROL
The flow control function is represented by the toolbar icon:
If a non-manual flow control option is specified in
[Run DUT]
mode or the reference molbox
has the MFC control option in
[Run molbox]
mode, this function is enabled to control the
flow or molbox output. In
[Run DUT]
mode, the flow can be changed by directly entering the
target flow, scrolling to the desired target flow, using the jog function or activating one of the
flow regulation modes. When the target flow is changed, an
<Activate Change>
button
appears for final confirmation prior to taking action. The
<DUT>
run screen will display the
resulting new set point when the target is accepted (see Section 6.7.3). Realize that the
target flow value is set in the target flow unit of measure. If these units are changed,
COMPASS
converts the current target flow value to the new unit.
The
[Tools], [Options], [Run Test] tab contains the overshoot tolerance that limits the
actual minimum and maximum set points based on the DUT’s range (see Section 11.6). To set
flows significantly above the DUT’s defined range, this value must be increased while
COMPASS is idle.
In
[Run molbox]
mode, there is no DUT information available to define the relationship
between analog output and flow. As a result, flow regulation and jog functions are not available.
The target flow field allows set points in volts and/or mA only based on the reference molbox.
Table 7 provides information on each of the flow control panel fields.
Press the <Activate Change> button to accept a new target flow value.
Figure 18.
<Flow Control>
Panel
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...