COMPASS
for molbox™ User’s Manual
Page 191
© 1998-2003
DH Instruments, Inc
.
19.6 TOLERANCE
The determination of whether a point is in or out of tolerance depends on the type of tolerance selected,
and the DUT and reference flows. Tolerance values may be specified in multiple units (see
Section 8.5.5).
COMPASS
first determines the tolerance value in DUT flow units and then compares the
value to the current DUT flow error to determine if the error is smaller (in) or larger (out) than
the tolerance. When the DUT tolerance selection is (%D %Reading) the tolerance is the sum of
the two values. If the selection is %DUTspan or %Reading, the tolerance used is the greater of the two
values. Tolerances that are defined in DUT flow, set, or output units are also compared directly in flow
units. The conversion from set and output units is described in Section 19.3.
DUTspan
DUTspan
DUTspan
Q
T
QT
×
=
100
%
%
QT
T
Q
reading
reading
f
%
%
Re
=
×
100
VARIABLE
DESCRIPTION
T
DUTspan
%
%DUTFS tolerance value defined in the DUT definition.
Q
DUTspan
DUT flow range span as determined by the difference between the
maximum and minimum flow output values in the DUT Definition.
DUTspan
QT
%
Maximum allowable( DUT – reference flow) deviation based on a
%DUTFS tolerance.
Q
f
Re
Current reference flow.
T
reading
%
%Reading tolerance value defined in the DUT definition.
QT
reading
%
Maximum allowable (DUT – reference flow) deviation based on a
%Reading tolerance.
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...