COMPASS
for molbox™ User’s Manual
Page 209
© 1998-2003
DH Instruments, Inc
.
Stability Limit
A limit expressed in units of flow per second (e.g., sccm/second). The stability limit is
used as the molbox
Ready/Not Ready
criterion
Ready (
<*>
)
if rate is less than stability
limit,
Not Ready
(
<
↑
>
or
<
↓
>
) if rate is greater than stability limit. See also
Rate
.
Surrogate Gas
A gas whose behavior, from the standpoint of a device under test (DUT), is similar to
the process gas for which the device is to be characterized and used. A surrogate gas
is often used in calibration and testing when the process gas cannot be used for safety
or cost reasons. A DUT gas conversion factor (“K” factor) is used to convert from the
surrogate gas to the process gas.
System
Configuration
A listing of the specific hardware items that
COMPASS
uses by default to perform
various data acquisition and control functions. The System Configuration is made up
of individual Hardware Definitions.
Tare
“Zeroing” of the molbox pressure transducers at the upstream or downstream pressure
to eliminate the zero error in the measurement of the differential pressure across a
molbloc. molbox1 and molbox RFM have automated taring routines and
COMPASS
can cause these to execute during a test. See the molbox Operation and Maintenance
Manual,
[Tare]
Section.
Test Definition
A database record which defines all the aspects of a test procedure that
COMPASS
can run on a DUT. Test Definitions are associated with DUTs and control the test
procedure when a test is run.
Test Gas
The gas flowing through the molbloc that is being used to run the test or calibration.
The test gas, from the standpoint of the device under test, could be a surrogate gas or
the process gas.
Tolerance
Same as DUT tolerance.
Upstream
Location of point A relative to point B in a flow system in which point A is at a different
location in the reverse direction of the flow. For example, the upstream molbloc
pressure port is upstream relative to the downstream pressure port because it is at a
different location from the downstream port in the opposite direction of the flow.
Volume Flow Units
Units of measure of volume flow (sometimes called “actual flow”). Do not confuse
volume flow units with volumetrically based mass flow units. See the molbox Operation
and Maintenance Manual,
[Unit]
Section, for an extensive discussion of flow units of
measure.
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...