5.1.2 EXAMPLE
#2
Assuming the molbox1 from Example 1 is set up and it has the MFC control option,
“How do I calibrate a DUT that outputs 0 to 5 V corresponding to 0 to 1 slm?”
This example demonstrates the setup requirements of a common analog output DUT. The only
setup required to use the molbox MFC control option is to make sure the molbox is set up
with the proper remote interface in a Reference Hardware Definition and select the molbox in
the System Configuration (see Section 10).
5.1.2.1
SETUP <VOLTAGE/CURRENT> OUTPUT TYPE
Use the
[Setup]
,
[DUT]
menu choice to open the DUT Definition Editor (see Section 8).
Create a new DUT Definition and set up the basic header information corresponding
to the DUT. In the
[Range]
tab (see Section 8.5.4), select the flow range units
from the flow units list and enter the flow min and max values; 0 and 1 respectively.
Check the
<Controller>
checkbox on the
<Set Range>
panel. Now type or
select “V” as the output and set units. Enter 0 and 5 as the min and max output
and set point values. DUT’s that use voltage or current signals to indicate flow
should use the
<Voltage/Current>
<
Output Type>
. This tells
COMPASS
to use
the instrument currently selected as the
<DUT Output DMM>
in the System
Configuration to measure the DUT flow. Select
<Voltage/Current>
as the DUT
<
Set Mode>
. This tells
COMPASS
to use the
<Set Point Power Supply>
selected in the Hardware Definition to control the DUT flow. Use the
<Save>
toolbar icon to save the DUT.
5.1.2.2 REFERENCE
SETUP
Since we have assumed the reference molbox in Example #1 is already properly
set up, we only need to make the necessary Hardware Definition selections to
use the MFC control option. Use the
[Setup]
,
[System]
menu option to display
the
<System Config>
form (see Section 10.2). First select the molbox as the
<Reference>
. This automatically adds a
<Reference molbox>
option to the
<Set Point Power Supply>
,
<Set Point DMM>
and
<DUT Output DMM>
lists.
Select
<Reference molbox>
in all lists. If it were desired to set with the molbox but
measure with a DMM, the DMM should be selected in the
<DUT Output DMM>
list.
5.1.2.3
RUN THE TEST
Use
[Run]
,
[Run Test]
to start a test sequence. Select the DUT just created and
the desired Test Definition when prompted. There is no interface selection
required when setting up the DUT interface during the test initialization.
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...