%Reading errors are undefined near 0 (zero) reference flows.
Set point errors are determined using the measured DUT set point whenever possible. If a measured
value is not available, the nominal value is used as the set point.
E
Q
Q
Q
DUTFS
DUT
f
DUTspan
%
Re
=
−
×
100
E
Q
Q
Q
Rdg
DUT
f
f
%
Re
Re
=
−
×
100
100
_
%
×
−
=
DUTspan
Set
ref
Set
DUTFS
Q
Q
Q
E
100
_
%
×
−
=
set
set
ref
Set
Rdg
Q
Q
Q
E
VARIABLE
DESCRIPTION
Q
DUT
Current DUT flow.
Q
f
Re
Current reference flow.
Q
DUTspan
DUT flow range span as determined by the difference between the
maximum and minimum flow output values in the DUT Definition.
E
DUTFS
%
DUT percent of full scale error.
E
Rdg
%
DUT percent of reading error.
Q
Se
DUT set point flow in DUT flow units.
Set
DUTFS
E
_
%
DUT percent of full scale set point error.
Set
Rdg
E
_
%
DUT percent of reading set point error.
19.5 DENSITY
CORRECTIONS
COMPASS
includes two gas density correction options to assist in testing of gas density dependent
devices under conditions different from their normal operating pressure and temperature conditions. The two
correction types are square root and proportional.
The gas density correction consider only gas density and no other pressure or temperature based DUT
corrections. For example, gas viscosity changes with pressure and temperature and this can have a
significant effect on small float rotameters but
COMPASS
does make a viscosity correction. Gravity has
an effect on rotameters but
COMPASS
has no gravity correction. Generally, corrections other than
density are not significant relative to the tolerances of these types of devices but, since they will not be
Содержание COMPASS FOR MOLBOX
Страница 1: ...1998 2003 DH Instruments Inc COMPASS for molbox Flow Calibration Software User s Manual...
Страница 16: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 2 N NO OT TE ES S...
Страница 20: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 64: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 50 N NO OT TE ES S...
Страница 94: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 80 N NO OT TE ES S...
Страница 111: ...COMPASS for molbox User s Manual Page 97 1998 2003 DH Instruments Inc Figure 57 Auxiliary Data Device Selector...
Страница 112: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 98 N NO OT TE ES S...
Страница 134: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 120 Figure 68 Hardware Definition Interface Tab...
Страница 150: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 136 N NO OT TE ES S...
Страница 166: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 152 N NO OT TE ES S...
Страница 172: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 158 N NO OT TE ES S...
Страница 192: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 178 N NO OT TE ES S...
Страница 194: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 180 N NO OT TE ES S...
Страница 200: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 186 N NO OT TE ES S...
Страница 216: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 202 N NO OT TE ES S...
Страница 220: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 206 N NO OT TE ES S...
Страница 224: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 210 N NO OT TE ES S...
Страница 226: ...COMPASS for molbox User s Manual 1998 2003 DH Instruments Inc Page 212 N NO OT TE ES S...