5 Operation
Operating Instructions DAC Universal
5.3 Reprocessing with the DAC Universal
båÖäáëÜ
15. Place the Basket lid on the counter or put it in the lid holder with the
basket pointing upwards.
16. Check the instruments visually to make sure they are dry.
17. Check the instruments visually to make sure they have been cleaned.
18. Notice the cycle number and check on your documentation (printer,
software, etc.) that the cycle was conducted properly: For
sterilization, "Sterilization, time, ok" must be indicated.
19. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
20. Document the cycle according to the work instructions applicable in
your practice.
21. Open the snap lock on the basket and take the small lid off the basket.
22. Remove the solid instruments from the Basket lid.
23. Place the small lid back on the basket.
The solid instruments can now be used again.
Sterilization of unwrapped instruments is not sufficient for invasive
procedures. After wrapping the instruments another sterilization must
be conducted in a class B sterilizer (e.g. DAC Premium).