7 Errors / Display messages
Sirona Dental A/S
7.1 Display messages and what they mean
Operating Instructions DAC Universal
90
Insufficient pressure
during the pressure test.
Make sure your compressor is switched on.
Check the seal on the lid. If it appears to have a leak,
please contact your authorized dealer.
Check the lid of the drain filter housing. If it appears
to have a leak, please contact your authorized dealer.
Check whether the pressure at the "Air" connection at
the back of the unit is 5-8 bar. If pressure falls
significantly during use, contact the compressor
supplier.
If the error occurs repeatedly, please contact your
authorized dealer.
91
Pressure drop detected
during pressure test.
Check the seal on the lid. If it appears to have a leak,
please contact your authorized dealer.
Remove any dirt on the lid sealing or at the rim of the
chamber that may be present.
Make sure that the lid is placed properly in the lid
holder.
If the error occurs repeatedly, please contact your
authorized dealer.
92
The power supply of the
DAC Universal was
interrupted or the "Up"-
Key was activated during
the cycle.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
93
The steam generator is
overheated.
Switch the DAC Universal off immediately by pulling
the power cord out of the socket.
Please contact your authorized dealer.
94
Oil consumption during
standby.
Disconnect the DAC Universal from the power supply
and wait 10 seconds.
Then reconnect the DAC Universal to the power
supply and restart.
Code
no.
Pictogram
Description
Corrective action