3 Product description
Operating Instructions DAC Universal
3.2 Controls and functional elements of the DAC Universal
båÖäáëÜ
3.2.2
Lid
The following lids can be used for reprocessing in the DAC Universal:
● Standard lid for reprocessing unwrapped rotating handpieces.
See "Functions with the Standard lid [ → 16]" for functions and
"Reprocessing with the Standard lid [ → 90]" for instructions.
The maximum load on the Standard lid is 6 instruments.
The permissible weight of each instrument is 250 g (for the PCD it is
265 g).
The permissible total weight of the instruments to be sterilized is
600 g.
● Basket lid for reprocessing unwrapped solid instruments.
See "Functions with the Basket lid [ → 18]" for functions and
"Reprocessing with the Basket lid [ → 94]" for instructions.
The maximum load on the Basket lid is 5 solid instruments.
The permissible weight of each instrument is 250 g.
The permissible total weight of the instruments to be sterilized is
600 g.
● Flex lid for cleaning and disinfection of unpacked syringes, scalers
and scalertips.
See "Functions with the Flex lid [ → 19]“ for functions and
"Reprocessing with the Flex lid [ → 100]“ for instructions.
The maximum load on the Flex lid is 6 instruments.
The permissible weight of each instrument is 250 g.
The permissible total weight of the instruments to be disinfected is
600 g.