5 Operation
Sirona Dental A/S
5.1 Regular activities
Operating Instructions DAC Universal
4. Fold the chemical indicator into a V-shape.
5. Put the chemical indicator into the PCD test device.
6. Twist the PCD test device clockwise to close.
7. Press the PCD test device onto the ISO/ INTRAmatic adapter until
you hear and feel a slight click.
8. Put instruments onto the other 5 adapters as usual.
9. Pull lightly on the instruments to ensure that they are firmly attached
to the adapters.
10. Start a standard cycle. See "Reprocessing with the Standard lid
[ → 90]" for details of the procedure.
IMPORTANT
Make sure the PCD test device is completely closed.
A proper check of the PCD test device is not ensured if it is partly open.
CAUTION
Risk of burns!
Allow the extremely hot instruments to cool down after the end of the
cycle.