5 Operation
Operating Instructions DAC Universal
5.1 Regular activities
båÖäáëÜ
5
Operation
5.1
Regular activities
The DAC Universal was installed by a service technician according to the
instructions in the "Installation [ → 40]" section.
See below for an overview of the activities that must be conducted during
the regular use of the DAC Universal.
5.1.1
Pre- and post-reprocessing of instruments
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Predisinfect directly at the treatment center.
3. Transport the instrument to the hygiene room in a suitable transport
container.
Activities and relevant chapter in the Operating Instructions
„Pre- and post-reprocessing of instruments [ → 47]“
„Checking the reprocessing result with an indicator / PCD test device
[ → 48]“
Replacing the lubrication concentrate, see "Adding/replacing the
lubrication concentrate [ → 54]"
„Checking oil supply to the chamber [ → 105]“
„Filling the water tank [ → 56]“
„Checking water supply to the chamber [ → 105]“
IMPORTANT
Please follow the regulations and guidelines applicable in your country
in your practice.
IMPORTANT
Please follow the instrument manufacturer's instructions when handling
the instruments.
IMPORTANT
The chuck of the instruments must be lubricated regularly after a
mechanical cleaning according to the requirements of the instrument
manufacturer.
IMPORTANT
Condition immediately, or at the latest, one hour after treatment.