100
66 15 574 D3472
D3472
.
201.20.01
.
02
03.2019
5 Operation
Sirona Dental A/S
5.3 Reprocessing with the DAC Universal
Operating Instructions DAC Universal
5.3.4
Reprocessing with the Flex lid
The Flex lid is used to clean and thermally disinfect ultrasonic scaler
handpieces, scaler tips, airflow nozzles and sprayvit nozzles.
Tip: If there are many instruments to be reprocessed or to ensure the
instruments will be available again quickly, we recommend using a
second Flex lid.
For a detailed procedure for attaching the instruments to the adapters,
please refer to the separate operating instructions for adapters.
5.3.4.1
Start cycle on the Flex lid
✔
All settings of the DAC Universal were preset according to "Regular
activities [ → 47]".
✔
The DAC Universal is switched on and ready for operation.
✔
The maximum number of 6 instruments are ready for reprocessing.
✔
The Flex lid is at hand and cooled down.
1. Remove the Flex lid from the holder on the DAC Universal.
2. Place the Flex lid on the counter or put it in the lid holder.
3. Plug each instrument onto the corresponding adapter on the Flex lid.
For more information on attaching the instruments or changing the
adapters see the separate operating instructions for adapters.
4. Attach a Class 5/6 chemical indicator in the indicator holder, "Flex" lid
in the groove on the center post of the Flex lid. See "Checking the
reprocessing result with an indicator / PCD test device [ → 48]" for
more information on using the chemical indicator.
5. Insert the Flex lid in the holder of the DAC Universal in such a way
that the instruments are pointing down.
IMPORTANT
The Flex lid is only for cleaning and high-level disinfection.
WARNING
The instruments must not be wrapped when using the Flex lid, as
cleaning and disinfection would then fail.
IMPORTANT
Do not start the DAC Universal without a lid inserted into the lid holder.
NOTICE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure there
are no objects in the chamber.
NOTICE
The DAC Universal and/or the instruments can be damaged if a lid is not
inserted correctly.