4 Installation
Sirona Dental A/S
4.1 Requirements of the location / Ambient conditions
Operating Instructions DAC Universal
4
Installation
4.1
Requirements of the location / Ambient conditions
Requirements
To find a suitable location for the DAC Universal, observe the following
requirements:
● Open, well-ventilated area
● Position in the unclean area directly adjacent to the clean area in the
hygiene room
● Level, heat-resistant countertop
● Close to a compressed air and power source
● Recommended minimum distance from the wall: 10 cm
● Recommended minimum height: 70 cm
If the DAC Universal needs to be switched off completely, the power plug
must be accessible.
Ambient conditions
The DAC Universal may be used only indoors under the following
ambient conditions:
● Relative humidity from 60 % to max. 80 %
● Up to an altitude of 2,000 m above sea level
● Room temperature between 16 °C and 26 °C (61 °F and 79 °F)
● Fluctuations in electrical supply should remain within ± 10 % of the
nominal voltage
● Transient overvoltage according to installation category II
● Air pollution level < 2 (ISO 8573-1:2010)
4.2
Installation of connections
CAUTION
Risk of crushing
The recommended minimum height for installing the DAC Universal is
70 cm, in order to prevent injuries when removing the lid.
IMPORTANT
Only technicians trained by the manufacturer with a training certificate
issued by the manufacturer are qualified to install, service, repair, and
maintain the DAC Universal.
IMPORTANT
Please note that the DAC Universal must be warmed up to room
temperature before being connected.