5 Operation
Sirona Dental A/S
5.2 Using the menu and display information
Operating Instructions DAC Universal
5.2
Using the menu and display information
The ready screen is the primary screen, which allows the user to get an
overview of the status of the machine. It is displayed when the autoclave
has been started and the automatic heating is completed.
A
The selected lid is shown on the left hand side of the ready screen.
The lids are listed as following:
• Standard lid
• Basket lid
• Flex lid
Important: Unused lids can be deactivated or removed from the unit by a service technician. These lids
are then no longer visible on the display.
B
The selected program is displayed at the top of the ready screen. The program is shown as
"temperature/holding time".
C
The current time is always displayed in the top right hand corner.
D
Warning messages are displayed in the right side of the ready screen. These warnings are as following:
“Service needed” appears after 2 years or 3.000 cycles.
“Low oil” appears when there’s insufficient oil quantity in the oil bottle.
“Low water” appears when there’s insufficient water quantity in the water tank.
“Bad water quality” appears when the water quality isn’t good enough.
E
The dark blue field of the ready screen indicates that the autoclave is ready to run a cycle. The Check
& Clean section can be entered by pressing the “Left”-Key. The menu can be entered by pressing the
“Right”-Key.