7 Errors / Display messages
Operating Instructions DAC Universal
7.1 Display messages and what they mean
båÖäáëÜ
60
The chamber
temperature and the
theoretical steam
temperature deviate from
each other more than 2K.
Please contact your authorized dealer.
61
The software has not
registered a change in
temperature in the steam
generator for 3 minutes.
Please contact your authorized dealer.
62
The temperature in the
chamber did not reach
110 °C / 230 °F.
If the water level is too low, fill demineralized or
distilled water (water conductivity < 3.0 µS/cm) into
the water tank.
If the DAC Universal is in a room with a temperature
< 16 °C / 61 °F, ensure that operating conditions are
fulfilled.
Press the "Up" and "Right"-Keys simultaneously to
reset and then restart the DAC Universal.
If the error occurs repeatedly, please contact your
authorized dealer.
64
The safety system was
activated.
Please contact your authorized dealer.
65
The safety system is
defective.
Please contact your authorized dealer.
66
The temperature was too
high during sterilization.
Please contact your authorized dealer.
Code
no.
Pictogram
Description
Corrective action