4 Installation
Sirona Dental A/S
4.3 Using the lubrication concentrate
Operating Instructions DAC Universal
4.2.4.2
Connecting the siphon
For connecting the drain connection with the siphon, a size 6/4 mm PTFE
tube must be used. The length of the PTFE tube must not exceed 3 m.
✔
You are wearing your protective gear.
✔
The PTFE tube is at hand.
1. Make sure that the drain system is clear and no counter pressure can
build up.
2. The DAC Universal can be connected directly to the drainage system.
4.3
Using the lubrication concentrate
The NitramOil #2 bottle is not yet mounted when the unit is delivered. The
bottle must be mounted and oil pumped through the hoses before the first
use of the DAC Universal.
For further information how to mount an oil bottle, see “Adding/replacing
the lubrication concentrate [ → 54]“.
4.4
Mounting adapters
Up to 6 instruments can be reprocessed simultaneously in the DAC
Universal. The instruments are connected to the lid using adapters.
These adapters are not yet mounted when the unit is delivered.
An overview of all adapters, a detailed procedure for installing the
adapters on the respective lids and a detailed procedure for installing the
instruments on the adapters can be found in the separate operating
instructions for adapters.