Sirona Dental A/S
5 Operation
Operating Instructions DAC Universal
5.2 Using the menu and display information
båÖäáëÜ
Manual water channels:
Water can be manually pumped through the bottom adapter in the
chamber by selecting this option. This can be used to ensure that the
DAC Universal is able to bring water to the internal water channels of the
Standard and Flex lids.
Please see "Checking water supply to the lids [ → 106]" for further
instructions.
Drain the chamber
Drains the chamber through the drain outlet. Select this function, if there’s
residue in the chamber that needs to be drained. An empty lid must be
inserted into the autoclave.
About 60 seconds after starting this operation the lid will reopen.
The chamber should then be empty. If not, the operation can be repeated.
Drain the water tank
Automatically empties the water tank through the chamber into the drain.
IMPORTANT
After using the manual water operation the chamber and connecting
hoses needs to be cleaned for water residue, see "Drain the chamber
[ → 88]".