5 Operation
Operating Instructions DAC Universal
5.1 Regular activities
båÖäáëÜ
2. Attach the indicator holder,Flex lid in the groove on the center post of
the Flex lid.
3. Start a cycle. See "Reprocessing with the Flex lid [ → 100]" for the
detailed procedure.
4. After the end of the cycle, check that the chemical indicator indicates
the disinfection test was passed.
5.1.2.2
Testing with a PCD process challenge device
The PCD (process challenge device) is used in combination with a
chemical indicator to simulate a steam penetration test and confirm
proper sterilization.
Checking using a PCD test device is necessary in the following cases:
● Before using a DAC Universal after initial installation
● After major repair work
● As part of routine weekly testing
Process
✔
The DAC Universal is switched on and ready for a cycle.
✔
Up to 5 instruments are ready for reprocessing in the DAC Universal,
see "Pre- and post-reprocessing of instruments [ → 47]".
1. Remove the Standard lid from the holder on the DAC Universal.
2. Place the Standard lid on the counter or put it in the lid holder.
3. First twist the PCD test device in the middle counterclockwise.
IMPORTANT
An ISO/INTRAmatic adapter is needed to mount a PCD test device to
the Standard lid.