5 Operation
Sirona Dental A/S
5.3 Reprocessing with the DAC Universal
Operating Instructions DAC Universal
A prompt to press down on the safety ring appears on the display.
8. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
The Standard lid of the DAC Universal closes as soon as the
safety ring is pressed.
The DAC Universal automatically runs the cleaning, lubrication
(unless deactivated), and sterilization cycle.
The display shows at any time which phase the DAC Universal is
in, see section "Display information [ → 59]“.
When the selected cycle is completed the display will indicate as
seen on the picture.
The lid moves about halfway out of the chamber. This indicates
that a cycle has been completed and the instruments have been
sterilized.
The cycle number is shown next to the completed cycle symbol.
9. Press the "Down"-Key.
The Standard lid will open completely.
IMPORTANT
If the Standard lid only holds SIROPure instruments, lubrication must be
deactivated:
➢ Press the "Right"-Key to enter the "Deactivate lubrication" screen.
➢ Press the "Down"-Key to deactivate lubrication or press the "Up"-
Key to abort the "Deactivate lubrication" process.
The "Start cycle" screen reappears.
If lubrication is deactivated, a warning symbol is displayed on the
right side of the display, indicating that lubrication is deactivated.
CAUTION
Risk of burns!
Never hold on to the lid or hold your head directly over the DAC
Universal when it opens, as steam will emerge from the device.
WARNING
After completion of the cycle, make sure that all instruments are still
properly attached to the adapters. If instruments have fallen into the
chamber during the cycle or if an instrument is no longer properly
attached after the end of a cycle, cleaning and sterilization have failed.
In this case, reprocessing of the instruments must be repeated.