5 Operation
Operating Instructions DAC Universal
5.2 Using the menu and display information
båÖäáëÜ
The error code is displayed in the top left hand corner;
Displays when the water quality is insufficient.
The water quality needs to be 3.0 μS/cm or better.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.
The error code is displayed in the top left hand corner;
Displays when water is needed.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.
The error code is displayed in the top left hand corner;
Displays when oil is needed.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.
The error code is displayed in the top left hand corner;
Appears when the air pressure is too high.
Check and see if pressurized air of 5-8 bars is connected to the air inlet.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.
The error code is displayed in the top left hand corner;
Appears when the air pressure is too low.
Check and see if pressurized air of 5-8 bars is connected to the air inlet.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.
The error code is displayed in the top left hand corner;
Low water.
Make sure the DAC Universal is supplied with water from the water
treatment system.
For further troubleshooting, see "Errors / Display messages [ → 122]".
Press the “Up”-Key to return.