6 Cleaning and maintenance
Sirona Dental A/S
6.3 Checking water supply to the lids
Operating Instructions DAC Universal
Water is pumped through the internal hoses.
3. Interrupt this process by pressing the "Down"-Key as soon as water
flows into the chamber through the 4 holes in the bottom of the
chamber.
6.3
Checking water supply to the lids
Frequency: weekly or every 50 cycles (whichever comes first).
In order to make sure that the DAC Universal is able to bring water to the
chamber and internal channels of the Standard and Flex lids correctly,
this operation needs to be tested.
✔
The DAC Universal is switched on and ready for operation.
✔
The lid of the DAC Universal is removed and there is an unobstructed
view into the chamber.
✔
You are wearing your protective gear.
✔
The Standard lid, Flex lid (if owned) and black lid extension piece for
water are at hand.
✔
The Standard lid, the Flex lid and the chamber are cooled down.
1. Attach the black lid extension piece for water to the bottom of the lid.
2. Place the other end of the black lid extension piece for water on the
bottom adapter in the bottom of the chamber.
3. Assure the lid is attached vertically on the bottom adapter, as shown.
4. From the "check water in chamber" option, enter the "check water,
bottom adapter“ option by pressing the “Right”-Key.
5. Press the "Down"-Key.
Water is pumped through the internal water hoses and channels
of the lid.
IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.
IMPORTANT
The water supply to the lid needs only to be checked for the Standard
and Flex lids, since these are the only ones with internal water channels.