108
66 15 574 D3472
D3472
.
201.20.01
.
02
03.2019
6 Cleaning and maintenance
Sirona Dental A/S
6.5 Drain the water tank
Operating Instructions DAC Universal
4. From the "check water, bottom adapter" option, press the "Down"-
Key.
Water is pumped through the internal channels of the lid.
5. When water is visible from underneath the lid’s top plate, observe that
water is flowing freely from the 6 different adapters.
6. Repress the "Down“-Key.
7. Remove the white extension piece for oil from the chamber and the
lid.
8. Repeat the process for all available Standard lids.
6.5
Drain the water tank
Frequency: weekly or every 50 cycles (whichever comes first).
Drain the water tank
✔
The DAC Universal is switched on and ready for operation.
✔
The Standard lid of the DAC Universal is removed and there is an
unobstructed view into the chamber.
✔
The drain tank has been emptied or the drain tank water level has
been checked (if using a drain tank).
✔
The Standard lid is at hand and cooled down.
1. Insert an empty Standard lid into the holder of the DAC Universal.
2. From the "check water, bottom adapter" option, press the "Right"-Key
to enter the "Turn off automatic water" option.
3. Press the "Down"-Key to turn off automatic water and enter the "Drain
the water tank" option.
4. Press the "Down"-Key to start the draining of the water tank.
The lid closes automatically.
IMPORTANT
If the water does not flow as described above, please contact your
authorized dealer.
IMPORTANT
If you have installed the DAC Universal with a permanent water
connection, please turn off the water, before you drain the water tank.
IMPORTANT
If using a drain tank, it must be checked that the drain tank will be able
to contain the drained-out water. If the drain tank is almost full, it must
be emptied before you drain the tank.
IMPORTANT
The "Turn off automatic water" option will only be available if automatic
water is turned on before entering the "Check & Clean" menu.